首页> 外文OA文献 >A feasibility study of multislice x-ray CT imaging of gel dosimeters using the “zero scan” method
【2h】

A feasibility study of multislice x-ray CT imaging of gel dosimeters using the “zero scan” method

机译:使用“零扫描”方法对凝胶剂量计进行多层X射线CT成像的可行性研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

This study extends the ‘zero scan’ method for CT imaging of polymer gel dosimeters to include multi-slice acquisitions. Multi slice CT images consisting of 24 slices of 1.2 mm thickness were acquired of an irradiated polymer gel dosimeter, and processed with the zero scan technique. The results demonstrate that zero scan based gel readout can be successfully applied to generate a three dimensional image of the irradiated gel field. Compared to the raw CT images the processed figures and cross gel profiles demonstrated reduced noise and clear visibility of the penumbral region. Moreover these improved results further highlight the suitability of this method in volumetric reconstruction with reduced CT data acquisition per slice. This work shows that 3D volumes of irradiated polymer gel dosimeters can be acquired and processed with x-ray CT.
机译:这项研究扩展了聚合物凝胶剂量计CT成像的“零扫描”方法,使其包括多层采集。用辐照的聚合物凝胶剂量计获取由24个1.2毫米厚的切片组成的多层CT图像,并用零扫描技术对其进行处理。结果表明,基于零扫描的凝胶读数可以成功地应用于生成辐射凝胶场的三维图像。与原始CT图像相比,处理后的图形和交叉凝胶曲线显示出噪声减少,半影区清晰可见。此外,这些改进的结果进一步强调了该方法在体积重建中的适用性,同时减少了每片CT数据的获取。这项工作表明可以用X射线CT采集和处理3D体积的辐照聚合物凝胶剂量计。