Technical note : modelling a complex micro-multileaf collimator using the standard BEAMnrc distribution

机译:技术说明:使用标准BEAMnrc分布对复杂的微多叶准直器建模

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose: The component modules in the standard BEAMnrc distribution may appear to be insufficient to model micro-multileaf collimators that have tri-faceted leaf ends and complex leaf profiles. This note indicates, however, that accurate Monte Carlo simulations of radiotherapy beams defined by a complex collimation device can be completed using BEAMnrc's standard VARMLC component module.----------ududMethods:udThat this simple collimator model can produce spatially and dosimetrically accurate micro-collimated fields is illustrated using comparisons with ion chamber and film measurements of the dose deposited by square and irregular fields incident on planar, homogeneous water phantoms.----------ududResults:udMonte Carlo dose calculations for on- and off-axis fields are shown to produce good agreement with experimental values, even upon close examination of the penumbrae.---------ududConclusions:udThe use of a VARMLC model of the micro-multileaf collimator, along with a commissioned model of the associated linear accelerator, is therefore recommended as an alternative to the development or use of in-house or third-party component modules for simulating stereotactic radiotherapy and radiosurgery treatments. Simulation parameters for the VARMLC model are provided which should allow other researchers to adapt and use this model to study clinical stereotactic radiotherapy treatments.
机译:目的:标准BEAMnrc分布中的组件模块似乎不足以对具有三面​​叶端和复杂叶轮廓的微多叶准直器建模。但是,此注释表明,可以使用BEAMnrc的标准VARMLC组件模块来完成由复杂的准直设备定义的放射治疗束的精确蒙特卡洛模拟。---------- ud udMethods: ud通过与离子室和薄膜的比较,说明该模型可以产生空间和剂量学上精确的微准直场,并通过对入射在平面均质水体上的正方形和不规则场沉积的剂量进行了测量。 ud结果: ud即使在仔细检查半影的情况下,对轴上和离轴场的蒙特卡洛剂量计算也显示出与实验值的良好一致性.--------- ud ud结论: ud因此,建议使用微型多叶准直仪的VARMLC模型以及相关线性加速器的调试模型,作为开发或使用内部或第三方组件模块来模拟立体定向雷达的替代方法放射疗法和放射外科治疗。提供了VARMLC模型的仿真参数,这些参数应允许其他研究人员改编并使用此模型来研究临床立体定向放射疗法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号