High precision, low disturbance calibration system for the CMS Barrel Electromagnetic Calorimeter High Voltage apparatus

机译:CMS桶式电磁热量计高压设备的高精度,低干扰校准系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The CMS Electromagnetic Calorimeter utilizes scintillation lead tungstate crystals, with avalanche photodiodes (APD) as photo-detectors in the barrel part. 1224 HV channels bias groups of 50 APD pairs, each at a voltage of about 380 V. The APD gain dependence on the voltage is 3%/V. A stability of better than 60 mV is needed to have negligible impact on the calorimeter energy resolution. Until 2015 manual calibrations were performed yearly. A new calibration system was deployed recently, which satisfies the requirement of low disturbance and high precision. The system is discussed in detail and first operational experience is presented.
机译:CMS电磁热量计利用闪烁的钨酸铅晶体,并在桶形部分中使用雪崩光电二极管(APD)作为光电探测器。 1224个HV通道对50个APD对的偏置组,每个对的电压约为380V。APD对电压的增益依赖性为3%/ V。需要一个高于60 mV的稳定性,以对量热仪的能量分辨率产生可忽略的影响。直到2015年,每年都要进行手动校准。最近部署了新的校准系统,可以满足低干扰和高精度的要求。对该系统进行了详细讨论,并提供了首次使用经验。

著录项

  • 作者

    Fasanella, Giuseppe;

  • 作者单位
  • 年度 2017
  • 总页数
  • 原文格式 PDF
  • 正文语种 en
  • 中图分类