Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements using the Eye Tracker and Retest Function of Spectralis(R) SD-OCT in Glaucomatous Eyes and Healthy Controls

机译:使用眼动仪测量视网膜神经纤维层厚度的可重复性以及青光眼和健康对照组的Spectralis(R)SD-OCT的复验功能

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

PURPOSE. To evaluate the impact of self-acting eyetracking and retest software on the reproducibility of retinal nerve fiber layer (RNFL) thickness measurements in glaucoma patients and healthy control subjects using Spectralis(®) SD-OCT. METHODS. RNFL thickness was measured in 56 normal and 47 glaucomatous eyes by one operator within one session with a brief rest between measurements. Three measurements were taken with eyetracker and retest function engaged (Method A), and three measurements were taken without eye tracker and without retest function (Method B). Measurements with Method A and Method B were taken alternately. RESULTS. Reliability, measured by intraclass correlation coefficient (ICC) for absolute agreement and coefficient of variation (COV), was calculated for the globalmean RNFL thickness (G), for each sector and for the peripapillaryboundle. The ICC (and lower 95% confidence interval (CI)) for the global mean RNFL thickness(G) for measurements using Method A in both normal and glaucomatous eyes was 0.99 (0.98 CI). In glaucomatous eyes the COV for measurements using Method B was between 2.7% and 10.5%, and between 1.3% and 3.5% formeasurements usingMethod A. CONCLUSION. Reproducibility of RNFL measurements with Spectralis(®) SD-OCT is excellent in bothnormal and glaucomatous eyes and can be significantly improved by using its eye trackerand retest software. The gain of reproducibility by using the software is significantly higher in glaucomatous eyes than in normal eyes. These findings suggest that software applications are capable ofsignificantly improving reproducibility of RNFL thickness measurements.
机译:目的。为了评估自作用眼动追踪和重新测试软件对使用Spectralis®SD-OCT在青光眼患者和健康对照组中视网膜神经纤维层(RNFL)厚度测量值的可重复性的影响。方法。一位操作员在一个疗程中对56只正常和47只青光眼眼进行RNFL厚度测量,并在测量之间短暂休息。在启用眼动仪和重新测试功能的情况下进行了三项测量(方法A),在没有眼动仪且没有重新测试功能的情况下进行了三项测量(方法B)。交替进行方法A和方法B的测量。结果。通过组内相关系数(ICC)的绝对一致性和变异系数(COV)来计算可靠性,计算的是全球平均RNFL厚度(G),每个扇区以及周围的乳突周围。使用方法A在正常和青光眼眼中进行测量的RNFL总体平均厚度(G)的ICC(以及较低的95%置信区间(CI))为0.99(0.98 CI)。在青光眼中,使用方法B测量的COV为2.7%至10.5%,使用方法A测量的COV为1.3%至3.5%。结论。使用Spectralis®SD-OCT进行RNFL测量的可重复性在正常和青光眼中均非常出色,并且可以通过使用其眼动仪和重新测试软件来显着改善。通过使用该软件,在青光眼眼中获得的重现性明显高于正常眼。这些发现表明,软件应用程序能够显着提高RNFL厚度测量的可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号