Performing gait analysis in a clinical setting can often be challenging due to time, cost and the availability of sophisticated three-dimensional (3D) gait analysis systems. This study has developed and tested a portable wireless gait assessment tool (wi-GAT) to address these challenges. Methods: Ten healthy volunteers participated in the study (age range 23–30 years). Spatio-temporal gait parameters were recorded simultaneously by the Vicon and the wi-GAT systems as each subject walked at their self-selected speed. Results: The stride length and duration, cadence, stance duration and walking speed recorded using the wi-GAT showed strong agreement with those same parameters recorded by the Vicon (ICC of 0.94–0.996). A difference between the systems in registering “toe off” resulted in less agreement (ICC of 0.299–0.847) in gait parameters such as %stance and %swing and DST. Discussion and conclusion: The study demonstrated good concurrent validity for the wi-GAT system. The wi-GAT has the potential to be a useful assessment tool for clinicians.
机译:由于时间,成本以及复杂的三维(3D)步态分析系统的可用性,在临床环境中进行步态分析通常可能具有挑战性。这项研究已经开发并测试了便携式无线步态评估工具(wi-GAT),以应对这些挑战。方法:十名健康志愿者参加了研究(年龄范围23-30岁)。当每个受试者以自己选择的速度行走时,Vicon和wi-GAT系统会同时记录时空步态参数。结果:使用wi-GAT记录的步幅和持续时间,步频,步态持续时间和步行速度与Vicon记录的相同参数具有很强的一致性(ICC为0.94-0.996)。系统在注册“趾距”方面的差异导致步态参数(如%stance和%swing和DST)的一致性较低(ICC为0.299-0.847)。讨论与结论:该研究证明了wi-GAT系统具有良好的并发有效性。 wi-GAT有可能成为临床医生有用的评估工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号