Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas

机译:基于Scheimpflug的非接触式设备在正常角膜上进行眼生物力学数据测量的可重复性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

PURPOSE\ud\udTo analyze the repeatability of a new device measuring ocular biomechanical properties, central corneal thickness (CCT), and intraocular pressure (IOP) and to investigate these parameters and their correlations in healthy eyes.\ud\udMETHODS\ud\udThree consecutive measurements were performed on each eye using the CorVis ST device (Oculus Optikgeräte, Inc., Wetzler, Germany). Ten specific parameters, CCT, and IOP were measured. Biometric data were recorded with IOLMaster (Carl Zeiss Meditec, Jena, Germany).\ud\udRESULTS\ud\udThis study comprised 75 eyes of 75 healthy volunteers (mean age: 61.24 ± 15.72 years). Mean IOP was 15.02 ± 2.90 mm Hg and mean CCT was 556.33 ± 33.13 μm. Intraclass correlation coefficient (ICC) was 0.865 for IOP and 0.970 for CCT, and coefficient of variation was 0.069 for IOP and 0.008 for CCT. ICC was 0.758 for maximum amplitude at highest concavity and 0.784 for first applanation time, and less than 0.6 for all other parameters. The device-specific data showed no significant relationship with age and axial length. Flattest and steepest keratometric values and IOP showed a significant correlation with the 10 device-specific parameters.\ud\udCONCLUSIONS\ud\udThe CorVis ST showed high repeatability for only IOP and pachymetric values. Single measurements are not reliable for the 10 device-specific parameters. The device allows for conducting clinical examinations and screening for surgeries altering ocular biomechanical properties with some form of averaging of multiple measurements.
机译:目的\ ud \ ud分析一种用于测量眼生物力学特性,中央角膜厚度(CCT)和眼内压(IOP)的新设备的可重复性,并研究健康眼睛中的这些参数及其相关性。\ ud \ udMETHODS \ ud \ ud三使用CorVis ST设备(德国Wetzler的OculusOptikgeräte公司)对每只眼睛进行连续测量。测量了十个特定参数CCT和IOP。用IOLMaster(卡尔·蔡司·梅迪泰克,耶拿,德国)记录生物数据。\ ud \ udRESULTS \ ud \ ud这项研究由75名健康志愿者(平均年龄:61.24±15.72岁)的75只眼睛组成。平均IOP为15.02±2.90 mm Hg,平均CCT为556.33±33.13μm。类内相关系数(ICC)IOP为0.865,CCT为0.970,变异系数IOP为0.069,CCT为0.008。最大凹度下最大振幅的ICC为0.758,第一次压平时ICC为0.784,所有其他参数的ICC小于0.6。特定于设备的数据显示与年龄和轴向长度没有显着关系。最平坦和最陡峭的角膜曲率值和IOP与10个特定于设备的参数显示显着相关。\ ud \ ud结论\ ud \ udCorVis ST仅对IOP和测厚法值显示出高重复性。对于10个设备特定的参数,单次测量不可靠。该设备允许进行临床检查和筛查,以改变多次测量的平均形式改变眼的生物力学特性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号