Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation

机译:Loop Recorder植入术后早期早期的不适当的心搏停止检测

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The implantable loop recorder is a useful diagnostic tool for patients with unexplained syncope. The capability to automatically detect and store arrhythmic events, implemented in the last generations of these devices, can further improve the diagnostic yield, but this feature can be compromised by inappropriate detection of false arrhythmias. We herein report the case of a patient in which several inappropriate activations of long-lasting asystole occurred in the two days following the implant, probably because of an intermittently loose contact between the device and subcutaneous tissue for a small pocket haematoma.
机译:植入式环路记录仪是用于原因不明的晕厥患者的有用诊断工具。在这些设备的最后几代中实现的自动检测和存储心律不齐事件的功能可以进一步提高诊断率,但是此功能可能会由于错误的错误心律失常检测而受到损害。我们在这里报道了一个患者的情况,其中在植入后的两天内发生了一些不适当的持久性心搏停止激活,这可能是由于该设备与小口袋血肿的皮下组织之间的间歇性松动。