The non-uniformity correction factor for the cylindrical ionization chambers in dosimetry of an HDR 192Ir brachytherapy source

机译:HDR 192Ir近距离放射治疗源剂量学中圆柱电离室的不均匀校正因子

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study is to derive the non-uniformity correction factor for the two therapy ionization chambers for the dose measurement near the brachytherapy source. The two ionization chambers of 0.6 cc and 0.1 cc volume were used. The measurement in air was performed for distances between 0.8 cm and 20 cm from the source in specially designed measurement jig. The non-uniformity correction factors were derived from the measured values. The experimentally derived factors were compared with the theoretically calculated non-uniformity correction factors and a close agreement was found between these two studies. The experimentally derived non-uniformity correction factor supports the anisotropic theory.
机译:这项研究的目的是为近距离放射治疗源附近的剂量测量推导两个治疗电离室的不均匀校正因子。使用两个体积分别为0.6 cc和0.1 cc的电离室。在特殊设计的测量夹具中,在距源源0.8 cm至20 cm的距离内进行空气中的测量。非均匀校正因子是从测量值得出的。将实验得出的因素与理论计算的非均匀性校正因素进行比较,发现这两项研究之间存在密切的一致性。实验得出的非均匀性校正因子支持各向异性理论。