Effect of surface waves on radiotherapy dosimetric measurements in water tanks

机译:表面波对水箱放射治疗剂量学测量的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The effect of surface waves, generated by moving the scanning arms in water phantoms, on radiation dosimetry is studied. It is shown that in large water tanks, high arm speeds can result in dosimetric errors of up to 5%. The measurements that are started after damping the water waves can result in about a 50% improvement in accuracy of measurements. It is shown that the water surfaces at the start of the measurements have high fluctuations that transform to a steady phase by elapsing time.
机译:研究了通过在水体模中移动扫描臂而产生的表面波对辐射剂量的影响。结果表明,在大型水箱中,高臂速会导致高达5%的剂量误差。在阻尼水波之后开始的测量可以使测量精度提高约50%。结果表明,测量开始时的水面波动很大,随着时间的流逝会转变为稳定的相位。

著录项

  • 作者

    Bakhtiari, Mohammad;

  • 作者单位
  • 年度 2011
  • 总页数
  • 原文格式 PDF
  • 正文语种 en