首页> 外文OA文献 >Light sensitometry of mammography films at varying development temperatures and times
【2h】

Light sensitometry of mammography films at varying development temperatures and times

机译:乳腺X射线摄影胶片在不同显影温度和时间下的光敏度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Kodak MinR-2000 mammography film is widely used for mammography imaging. The sensitometric indices like base plus fog level (B + F), maximum optical density (ODmax), average gradient (AG) and speed of this film at varying development temperatures and times were evaluated using a light sensitometer. Totally 33 film strips were cut from a single Kodak MinR-2000 mammography film box and exposed in a light sensitometer operated in the green light spectrum to produce a 21-step sensitometric strip. These exposed film strips were processed at temperatures in the range of 32°C–37°C in the step of 1°C and at processing times in the range of 1–6 minutes in the step of 1 minute. The results of the present study show that the measured base plus fog level of the mammography film was not affected much, whereas significant changes were seen in the ODmax, AG and speed with varying development temperatures and times. The ODmax values of the film were found in the range of 3.67–3.76, AG values were in the range of 2.48–3.4 and speed values were in the range of 0.015–0.0236 when the processing temperature was varied from 32°C to 37°C. With processing time variation from 1 to 6 minutes, the observed changes in ODmax values were in the range of 3.54-3.71, changes in AG were in the range of 2.66–3.27 and changes in speed were in the range of 0.011–0.025. Based on these observations, recommendations for optimum processing parameters to be used for this film are made.
机译:柯达MinR-2000乳腺X线摄影胶片广泛用于乳腺X线摄影成像。使用光敏计评估了感光度指数,如碱加雾度(B + F),最大光密度(ODmax),平均梯度(AG)和该膜在变化的显影温度和时间下的速度。从单个Kodak MinR-2000乳腺摄影胶片盒中切下总共33条胶片,并在以绿色光谱操作的光敏计中曝光,以制作21步的感光带。这些曝光的胶片条在1°C的步骤中于32°C–37°C的温度范围内处理,在1分钟的步骤中在1-6分钟的范围内处理时间。本研究的结果表明,乳腺X线摄影胶片的底基和雾度没有受到太大的影响,但是随着显影温度和显影时间的变化,ODmax,AG和速度也发生了显着变化。当处理温度从32°C更改为37°C时,膜的ODmax值在3.67–3.76范围内,AG值在2.48–3.4范围内,速度值在0.015–0.0236范围内C。随着处理时间从1分钟到6分钟的变化,观察到的ODmax值的变化在3.54-3.71范围内,AG的变化在2.66-3.27范围内,速度的变化在0.011-0.025范围内。基于这些观察,提出了用于该膜的最佳加工参数的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号