Measurement of Blood Pressure Using an Arterial Pulsimeter Equipped with a Hall Device

机译:使用配有霍尔装置的动脉脉搏计测量血压

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

To measure precise blood pressure (BP) and pulse rate without using a cuff, we have developed an arterial pulsimeter consisting of a small, portable apparatus incorporating a Hall device. Regression analysis of the pulse wave measured during testing of the arterial pulsimeter was conducted using two equations of the BP algorithm. The estimated values of BP obtained by the cuffless arterial pulsimeter over 5 s were compared with values obtained using electronic or liquid mercury BP meters. The standard deviation between the estimated values and the measured values for systolic and diastolic BP were 8.3 and 4.9, respectively, which are close to the range of values of the BP International Standard. Detailed analysis of the pulse wave measured by the cuffless radial artery pulsimeter by detecting changes in the magnetic field can be used to develop a new diagnostic algorithm for BP, which can be applied to new medical apparatus such as the radial artery pulsimeter.
机译:为了在不使用袖带的情况下测量精确的血压(BP)和脉搏率,我们开发了一种动脉脉搏计,该脉搏计由装有霍尔装置的小型便携式设备组成。使用两个BP算法方程对在动脉搏动计测试期间测得的脉搏波进行回归分析。将通过无袖动脉搏动计在5 s内获得的BP估计值与使用电子或液体汞BP计获得的值进行比较。收缩压和舒张压BP的估计值与测量值之间的标准差分别为8.3和4.9,接近BP国际标准的值范围。通过对袖带式radial动脉脉搏计测量磁场的变化进行详细分析,可用于开发一种新的BP诊断算法,该算法可应用于new动脉脉搏计等新型医疗器械。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号