首页> 外文OA文献 >Placement of the Internal Pulse Generator for Deep Brain Stimulation in the Upper Back to Prevent Fracture of the Extension Wire due to Generator Rotation: Case Report
【2h】

Placement of the Internal Pulse Generator for Deep Brain Stimulation in the Upper Back to Prevent Fracture of the Extension Wire due to Generator Rotation: Case Report

机译:内部脉冲发生器用于在上背部深层大脑刺激,以防止由于发生器旋转而引起的延长线骨折:病例报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Deep brain stimulation (DBS) is a common surgical procedure used for the treatment of Parkinson's disease (PD) and essential tremor. A potential complication of this procedure is hardware failure. The authors report a case of DBS hardware failure in which repeated fractures of the extension wire were caused by abnormal rotational movements of the IPG placed in the loose subclavicular tissue of an overweight female. Implantation of the IPG in the suprascapular area prevented further extension wire fractures. This strategy may be especially relevant in overweight females with loose subclavicular tissue.
机译:深部脑刺激(DBS)是用于治疗帕金森氏病(PD)和原发性震颤的常见外科手术方法。该过程的潜在复杂性是硬件故障。作者报告了一例DBS硬件故障,其中延长线反复断裂是由放置在超重女性锁骨下松弛组织中的IPG异常旋转运动引起的。将IPG植入肩area上区域可防止进一步的延长线骨折。这种策略在锁骨下组织松弛的超重女性中可能特别重要。