This paper explores the combination of cone beam optical computed tomography with an N-isopropylacrylamide (NIPAM)-based polymer gel dosimeter for three-dimensional dose imaging of small field deliveries. Initial investigations indicate that cone beam optical imaging of polymer gels is complicated by scattered stray light perturbation. This can lead to significant dosimetry failures in comparison to dose readout by magnetic resonance imaging (MRI). For example, only 60% of the voxels from an optical CT dose readout of a 1 l dosimeter passed a two-dimensional Low's gamma test (at a 3%, 3 mm criteria, relative to a treatment plan for a well-characterized pencil beam delivery). When the same dosimeter was probed by MRI, a 93% pass rate was observed. The optical dose measurement was improved after modifications to the dosimeter preparation, matching its performance with the imaging capabilities of the scanner. With the new dosimeter preparation, 99.7% of the optical CT voxels passed a Low's gamma test at the 3%, 3 mm criteria and 92.7% at a 2%, 2 mm criteria. The fitted interjar dose responses of a small sample set of modified dosimeters prepared (a) from the same gel batch and (b) from different gel batches prepared on the same day were found to be in agreement to within 3.6% and 3.8%, respectively, over the full dose range. Without drawing any statistical conclusions, this experiment gives a preliminary indication that intrabatch or interbatch NIPAM dosimeters prepared on the same day should be suitable for dose sensitivity calibration.
机译:本文探讨了锥形束光学计算机断层扫描技术与基于N-异丙基丙烯酰胺(NIPAM)的聚合物凝胶剂量计的结合,用于小剂量野外给药的三维剂量成像。初步研究表明,聚合物凝胶的锥形束光学成像会因散射杂散光的干扰而变得复杂。与通过磁共振成像(MRI)进行剂量读取相比,这可能导致明显的剂量测定失败。例如,从1 l剂量计的光学CT剂量读数中,只有60%的体素通过了二维Low'sγ测试(相对于特征明确的笔形射束的治疗计划,其标准为3%,3 mm)交货)。通过MRI探测相同的剂量计时,观察到93%的通过率。修改剂量计制剂后,光学剂量测量得到了改善,使其性能与扫描仪的成像能力相匹配。使用新的剂量计准备,99.7%的光学CT体素在3%,3毫米标准下通过了Low's伽玛测试,在2%,2mm标准下通过了92.7%。发现(a)同一批凝胶和同一天(b)不同批制备的改良剂量计的小样品集的罐内剂量响应分别在3.6%和3.8%之内一致。在整个剂量范围内。在没有得出任何统计结论的情况下,该实验初步表明,同一天准备的批内或批间NIPAM剂量计应适用于剂量敏感性校准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号