首页> 外文OA文献 >A Fluorescence-Based Technique to Construct Size Distributions from Single-Object Measurements: Application to the Extrusion of Lipid Vesicles
【2h】

A Fluorescence-Based Technique to Construct Size Distributions from Single-Object Measurements: Application to the Extrusion of Lipid Vesicles

机译:基于荧光的单对象测量构建尺寸分布的技术:在脂质囊泡挤出中的应用

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We report a novel approach to quantitatively determine complete size distributions of surface-bound objects using fluorescence microscopy. We measure the integrated intensity of single particles and relate it to their size by taking into account the object geometry and the illumination profile of the microscope, here a confocal laser scanning microscope. Polydisperse (as well as monodisperse) size distributions containing objects both below and above the optical resolution of the microscope are recorded and analyzed. The data is collected online within minutes, which allows the user to correlate the size of an object with the response from any given fluorescence-based biochemical assay. We measured the mean diameter of extruded fluorescently labeled lipid vesicles using the proposed method, dynamic light scattering, and cryogenic transmission electron microscopy. The three techniques were in excellent agreement, measuring the same values within 7–9%. Furthermore we demonstrated here, for the first time that we know of, the ability to determine the full size distribution of polydisperse samples of nonextruded lipid vesicles. Knowledge of the vesicle size distribution before and after extrusion allowed us to propose an empirical model to account for the effect of extrusion on the complete size distribution of vesicle samples.
机译:我们报告了一种新颖的方法,可以定量确定使用荧光显微镜检测的表面结合物体的完整尺寸分布。我们通过考虑物体的几何形状和显微镜(这里是共聚焦激光扫描显微镜)的照明轮廓来测量单个粒子的集成强度,并将其与它们的大小相关联。记录并分析包含低于和低于显微镜光学分辨率的物体的多分散(以及单分散)尺寸分布。几分钟之内就可以在线收集数据,这使用户可以将对象的大小与任何基于荧光的生化分析的响应相关联。我们使用提出的方法,动态光散射和低温透射电子显微镜测量了挤出的荧光标记脂质囊泡的平均直径。三种技术非常吻合,在7–9%的范围内测量相同的值。此外,我们在这里首次证明了确定非挤出脂质囊泡多分散样品全尺寸分布的能力。挤压前后囊泡大小分布的知识使我们能够提出经验模型,以说明挤压对囊泡样品完整大小分布的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号