首页> 外文OA文献 >A fully automated flow injection atomic absorption system for the determination of copper traces in waters with on-line pre-concentration in an ion-exchange column
【2h】

A fully automated flow injection atomic absorption system for the determination of copper traces in waters with on-line pre-concentration in an ion-exchange column

机译:全自动流动注射原子吸收系统,用于测定 离子交换柱在线富集测定水中的痕量铜

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The paper describes the development of an automatic on-line columnpre-concentration technique using a time based-flow injection atomic absorption spectrometry system. A manifold incorporating a micro-column containing 25 mg of Dowex 50W-X8 was used with a time-based injector for the pre-concentration and determination ofcopper in natural and drinking waters. The system features depend on the alternate positions of a solenoid valve. The 3σ detection limits, enrichment factors, sampling frequency, relative standard deviations and linear calibration graphs were, respectively, in the range 0.6-1.5 μg/l, 25-60, 15-30 measurements/h, 1.0-3.1% and 1-65 μg/ml for pre-concentration times of 1 min. The procedure was successfully applied to a range of water samples and the accuracy was assessed through recovery experiments, the analysis of certified reference water samples and by independent analysis byatomic absorption spectrometry with electrothermal atomization.
机译:本文描述了使用基于时间的流动注射原子吸收光谱系统的自动在线柱前浓缩技术的发展。将装有包含25 mg Dowex 50W-X8的微柱的歧管与基于时间的注射器一起使用,用于自然水和饮用水中铜的预浓缩和测定。系统功能取决于电磁阀的备用位置。 3σ的检出限,富集系数,采样频率,相对标准偏差和线性校正图分别在0.6-1.5μg/ l,25-60、15-30测量/h、1.0-3.1%和1- 65μg/ ml的1分钟预浓缩时间。该程序已成功应用于一系列水样,并通过回收实验,认证的参考水样分析以及通过电热雾化原子吸收光谱法进行独立分析来评估其准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号