首页> 外文OA文献 >The Adequacy of the Air-Mix Control in Ventilators Operated from an Oxygen Source
【2h】

The Adequacy of the Air-Mix Control in Ventilators Operated from an Oxygen Source

机译:用氧气源操作的通风机中空气混合控制的适当性

摘要

Mean inspired oxygen levels, delivered by the Bennett PR1 and Bird Mark 7 ventilators, were measured at various cycling pressures. Using the Bird ventilator, oxygen levels varied from 51.5 to 96.8%, being highest at low flows and at high cycling pressures. The Bennett PR1 delivers more constant percentages of oxygen, in the range 61.5 to 72.3%. With both makes of machine, each ventilator provides different inspired oxygen levels under comparable circumstances. It is recommended that this type of equipment be driven by compressed air, oxygen being added as required.
机译:在不同的循环压力下,对Bennett PR1和Bird Mark 7呼吸机提供的平均吸入氧气水平进行了测量。使用伯德呼吸机时,氧气含量从51.5至96.8%不等,在低流量和高循环压力下最高。 Bennett PR1提供更恒定的氧气百分比,范围为61.5至72.3%。对于这两种机器,在可比较的情况下,每个呼吸机都提供不同的吸入氧气水平。建议使用压缩空气驱动此类设备,并根据需要添加氧气。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号