首页> 外文OA文献 >An evaluation of the NR-1000F Auto Refractometer in high refractive errors.
【2h】

An evaluation of the NR-1000F Auto Refractometer in high refractive errors.

机译:NR-1000F自动折光仪在高折光误差方面的评估。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The manifest refractions of 37 high myopic and 16 high hypermetropic eyes in a total of 28 patients, aged 5 to 42 years, were determined on the Nikon NR-1000F Auto Refractometer. The data obtained were compared with the final clinical prescriptions for these eyes and analysed for degree of agreement for the spherical equivalents, sphere components, and cylinder powers and axes. The Auto Refractometer recordings for all the different refractive components were observed to skew towards more minus in the high myopic eyes but more plus in the high hypermetropes. The possible significance of these skewing patterns is discussed with reference to our earlier studies.
机译:在尼康NR-1000F自动验光仪上确定了28例年龄在5至42岁的28例患者中37眼高度近视眼和16眼高度远视眼的明显屈光度。将获得的数据与这些眼睛的最终临床处方进行比较,并分析球面等效物,球面组件以及圆柱度和轴的一致性程度。观察到所有不同屈光分量的自动折光仪记录在高度近视眼中偏向负号,而在高度远视眼中偏向正号。参照我们之前的研究讨论了这些倾斜模式的可能意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号