Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensity-modulated radiotherapy

机译:在小型监控器单元和小型段下执行prim-and-shoot强度调制放射治疗的西门子primus线性加速器的性能表征

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20×20 cm2 field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm.
机译:逐步调制强度调制放射疗法(IMRT)的实现需要仔细了解加速器的启动特性,以确保向患者准确而精确地传输放射剂量。在小MU的条件下,研究了西门子Primus线性加速器(LA)的剂量特性,该线性加速器分别以每分钟(min)300和500监视单位(MU)的剂量率传送6和18 MV X射线。 1至100。通过使用0.6 cc电离室在固体水体模中10 cm深度处测量电离,研究了在不同剂量校准参数(D1_C0)下的剂量监测器线性。从不同强度调制(IM)组研究了监控器单元的稳定性,该组包括每场和场数的MU的各种组合。使用2D Profiler等中心保持在5 cm的深度,在小MU下在20×20 cm2场下在正常模式和IMRT模式下研究了光束平坦度和对称性的稳定性。通过测量来自不同IM组的每MU的剂量,研究了1到10 MU之间的节间反应,每个IM组由四个节组成,节间间隔为2 cm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号