首页> 外文OA文献 >The Bimetric cementless total hip replacement: 7–18 year follow-up assessing the influence of acetabular design on survivorship
【2h】

The Bimetric cementless total hip replacement: 7–18 year follow-up assessing the influence of acetabular design on survivorship

机译:Bimetric非骨水泥全髋关节置换术:7-18年的随访评估髋臼设计对生存率的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study was to evaluate the mid- to long-term survivorship of Bimetric cementless total hip replacement and assess how it is affected by the acetabular design. This was a retrospective analysis of 127 Bimetric cementless total hip replacements in 110 patients with a follow-up of 7–18 years. A single design stem and three different cementless metal-backed acetabular designs were used. Patients were assessed clinically using the Harris hip score and radiologically by independent review of current hip radiographs. There was only one case of aseptic loosening of the femoral stem. The earliest acetabular design showed a high failure rate whilst the latter two designs showed a 96% survivorship at a mean of 9.5 years. We conclude that a combination of the bimetric stem with either of the latter acetabular cup designs has a good mid- to long-term performance.
机译:这项研究的目的是评估Bimetric非骨水泥全髋关节置换术的中长期存活率,并评估其如何受到髋臼设计影响。这是对110例患者的127例Bimetric非骨水泥全髋关节置换术的回顾性分析,随访7–18年。使用单个设计的茎和三个不同的非骨水泥金属支持的髋臼设计。使用Harris髋关节评分对患者进行临床评估,并通过对当前髋部X射线照片进行独立审查进行放射学评估。仅有一例股骨柄无菌性松动。最早的髋臼设计显示出较高的失败率,而后两种设计显示平均9.5年的生存率达96%。我们得出的结论是,双侧柄与后两种髋臼杯设计的组合具有良好的中长期性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号