首页> 外文OA文献 >Automated Detection of Malaria-Associated Pseudoeosinophilia and Abnormal WBC Scattergram by the Sysmex XE-2100 Hematology Analyzer: A Clinical Study with 1,801 Patients and Real-Time Quantitative PCR Analysis in Vivax Malaria-Endemic Area
【2h】

Automated Detection of Malaria-Associated Pseudoeosinophilia and Abnormal WBC Scattergram by the Sysmex XE-2100 Hematology Analyzer: A Clinical Study with 1,801 Patients and Real-Time Quantitative PCR Analysis in Vivax Malaria-Endemic Area

机译:Sysmex XE-2100血液分析仪自动检测与疟疾有关的假性嗜酸性粒细胞增多和WBC散点图异常:在Vivax疟疾流行地区对1,801名患者进行的临床研究和实时定量PCR分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Recently, the XE-2100 hematology analyzer was investigated in a rather small patient group; pseudoeosinophilia or abnormal white blood cell (WBC) scattergrams reported by this instrument were considered as significantly valuable diagnostic parameters in detecting vivax malaria. This study was conducted not only to assess the usefulness of pseudoeosinophilia or abnormal WBC scattergram in vivax malaria-endemic areas with large patient groups (N = 1,801) but also to investigate the correlation of parasitemia and platelet count with pseudoeosinophilia and abnormal WBC scattergrams. Of the 1,801 analyzed patients, 413 (22.9%) were found to have malaria by Wright–Giemsa stained blood smears. Overall, either pseudoeosinophilia or abnormal WBC scattergram was detected in 191 of 413 malaria patients and 4 of 1,388 patients without malaria (sensitivity = 46.2%, specificity = 99.7%). We suggest that clinical hematology laboratories using the XE-2100 analyzer should be aware of such specific parameters, even with the absence of a clinical request.
机译:最近,XE-2100血液分析仪在一个很小的患者组中进行了研究。该仪器报告的假嗜酸性粒细胞增多或异常白细胞(WBC)散点图被认为是检测间日疟的重要有价值的诊断参数。这项研究不仅是为了评估假性嗜酸性粒细胞增多或异常WBC散点图在有大量患者人群的间日疟疾流行地区(N = 1,801)的有用性,还调查了寄生虫血症和血小板计数与假性嗜酸粒细胞增多和异常WBC散点图的相关性。在Wright–Giemsa染色的血液涂片检查中,在1,801名分析的患者中,发现413名(22.9%)患有疟疾。总体而言,在413名疟疾患者中的191名和无疟疾的1,388名患者中有4名检测到伪嗜酸性粒细胞增多或WBC散点图异常(敏感性= 46.2%,特异性= 99.7%)。我们建议即使没有临床要求,使用XE-2100分析仪的临床血液学实验室也应了解此类特定参数。