X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation.
机译:X射线计算机断层扫描(CT)先前已被报道为聚合物凝胶(PAG)剂量计的评估工具。在这项研究中,优化了Siemens Emotion X射线CT扫描仪的成像协议,以评估PAGAT常氧凝胶剂量计。扫描参数被优化为130 kV和150 mA,对于较小的视野,切片厚度为3 mm,对于较大的照射区域,切片厚度为5 mm。将噪声降低到可接受水平的平均图像数量得出的结论是25。得出的结论是,使用10 mM THP所需的总单体浓度为7%,以达到剂量水平最高至CT值的最大变化X射线CT评估需要15 Gy。最佳扫描参数可能会因X射线CT扫描仪而异。因此,用于评估聚合物凝胶的每个扫描仪的成像规程需要针对凝胶剂量学评估的目的进行单独优化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号