Technical note: Heterogeneity dose calculation accuracy in IMRT: Study of five commercial treatment planning systems using an anthropomorphic thorax phantom

机译:技术说明:IMRT中的异质剂量计算准确性:使用拟人化的胸部模型对五个商业治疗计划系统的研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study was to determine the accuracy of five commonly used intensity-modulated radiation therapy (IMRT) treatment planning systems (TPSs), 3 using convolution superposition algorithms or the analytical anisotropic algorithm (CSA∕AAAs) and 2 using pencil beam algorithms (PBAs), in calculating the absorbed dose within a low-density, heterogeneous region when compared with measurements made in an anthropomorphic thorax phantom. The dose predicted in the target center met the test criteria (5% of the dose normalization point or 3 mm distance to agreement) for all TPSs tested; however, at the tumor-lung interface and at the peripheral lung in the vicinity of the tumor, the CSA∕AAAs performed better than the PBAs (85% and 50%, respectively, of pixels meeting the 5%∕3-mm test criteria), and thus should be used to determine dose in heterogeneous regions.
机译:这项研究的目的是确定五个常用的强度调制放射治疗(IMRT)治疗计划系统(TPS),三个使用卷积叠加算法或分析各向异性算法(CSA ∕ AAAs)以及两个使用笔形射束算法的准确性(PBA),与拟人化的胸模体中进行的测量相比,在计算低密度异质区域内的吸收剂量时。对于所有测试的TPS,目标中心预测的剂量均符合测试标准(剂量归一化点的5%或相距3 mm的距离);然而,在肿瘤-肺界面和肿瘤附近的周围肺部,CSA ∕ AAA的性能优于PBA(分别达到5%∕ 3-mm测试标准的像素分别为85%和50%) ),因此应用于确定异质区域的剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号