The accuracy and working condition of 210 sphygmomanometers were tested: 100 (50 and mercury and 50 aneroid) models were used in family practices and 100 mercury models in hospitals. Faults in the inflation-deflation system were common and caused mainly by dirt or wear in the control valves. Leakage occurred in 48% of the hospital and 33% of the family practice sphygmomanometers. In the mercury models the mercury or air vents were often in an unsatisfactory condition or the calibrated glass tube dirty. The accuracy of the gauges was examined at 90 and 150 mm Hg: fewer than 2% of the mercury sphygmomanometers but 30% of the aneroid models had errors greater than +/- 4 mm Hg at either pressure. Over half of the cuffs examined had bladders widths less than the recommended size, and 94% had bladders shorter than the length recommended for use on normal adults. Mercury sphygmomanometers should be bought in preference to aneroid models as they are more accurate, less expensive in the long term, and can be maintained by the owner; they should be checked every six to 12 months depending on usage. Replacement parts should be kept readily available.
机译:测试了210台血压计的准确性和工作条件:在家庭实践中使用了100种(50种汞和50种汞和50种无液)模型,在医院中使用了100种汞模型。充气-放气系统中的故障很常见,主要是由控制阀中的灰尘或磨损引起的。 48%的医院和33%的家庭实践血压计发生泄漏。在水银模型中,水银或通风孔经常处于不令人满意的状态,或者校准的玻璃管脏了。在90和150 mm Hg的情况下检查了压力计的精度:少于2%的水银血压计,但30%的无液模型在任一压力下的误差均大于+/- 4 mm Hg。所检查的袖带中,有一半以上的膀胱宽度小于建议的大小,而94%的膀胱长度小于建议用于正常成年人的长度。汞血压计应优先购买无液血压计,因为它们更准确,长期而言更便宜,并且可由所有者维护;根据使用情况,应每6至12个月检查一次。更换零件应保持随时可用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号