Equipment design issues for the in vivo X-ray fluorescence analysis of bone lead.

机译:骨铅的体内X射线荧光分析的设备设计问题。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Several groups have reported the development of systems, based on the principle of X-ray fluorescence, for the in vivo measurement of bone lead concentrations. These systems have used the detection of either the characteristic L or K X-rays resulting from excitation by a suitable photon source. This paper examines design issues related to the development of these systems. These design issues are, in most instances, a result of consideration of the physical principles involved, and hence there are many features common to the systems developed by the individual groups. Design issues discussed in this paper include the selection of the site for measurement, source-sample-detector configuration, and collimation. Specific examples from published work are used to demonstrate the relevant features.
机译:几个小组已经报告了基于X射线荧光原理的系统开发,用于体内测量骨铅浓度。这些系统已经使用了由合适的光子源激发产生的特征性L或K X射线的检测。本文研究了与这些系统的开发有关的设计问题。在大多数情况下,这些设计问题是考虑到所涉及物理原理的结果,因此,各个小组开发的系统具有许多共同的特征。本文讨论的设计问题包括测量地点的选择,源样本检测器配置和准直。已发表作品的特定示例用于演示相关功能。