A double-beam rapid-wavelength-scanning stopped-flow spectrophotometer system based on the Norcon model 501 spectrometer was construced, which enables u.v.-or visible absorbance spectra to be recorded at the rate of 800/s after the rapid mixing (within 3ms) of two reactant solutions. Each spectrum spans about 200nm in 1ms. It is possible to record difference spectra during reactions with half-lives less than 10ms involving absorbance changes of less than 0.1 absorbance unit. Analogue circuitry is used to produce spectra of absorbance against wavelength. Up to 32 such spectra can be recorded at pre-selected times during a reaction and stored in an 8Kx8-bit-word hard-wired data-capture system to be subsequently displaned individually or simultaneously. Time-courses at different wavelengths can also be displayed. By averaging up to 216 spectra it is possible to record spectra under conditions of low signal-to-noise ratios...
机译:构造了基于Norcon 501型光谱仪的双光束快速波长扫描停止流式分光光度计系统,该系统可在快速混合后(3毫秒内)以800 / s的速率记录紫外或可见吸收光谱。两种反应物溶液。每个光谱在1ms内跨越约200nm。可以记录半衰期小于10毫秒的反应过程中的差异光谱,其中半衰期小于0.1吸光度单位。模拟电路用于产生相对于波长的吸收光谱。在反应过程中,可以在预先选择的时间记录多达32个这样的光谱,并将其存储在8Kx8位字的硬连线数据捕获系统中,以便随后分别或同时进行分解。也可以显示不同波长的时程。通过平均多达216个光谱,可以在低信噪比的条件下记录光谱...

著录项

  • 作者

    Holloway, M R; White, H A;

  • 作者单位
  • 年度 1975
  • 总页数
  • 原文格式 PDF
  • 正文语种 en
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号