A portable indirect calorimeter adapted from adult use was validated for use in preterm infants. Oxygen consumption (VO2) and carbon dioxide production (VCO2) were subsequently measured in 16 preterm infants breathing spontaneously in room air (canopy mode) and in nine preterm infants receiving intermittent positive pressure ventilation (ventilator mode). Validation of the system was performed using a gas injection technique with nitrogen to simulate VO2 and carbon dioxide for VCO2. Mean errors in validation of the canopy mode were 1.4% and 0.2% for VO2 and VCO2 with limits of agreement of 0.6 (+2SD) ml/min and -1.3 (-2SD) ml/min, and 0.9 (+2SD) ml/min and -2.3 (-2SD) ml/min respectively. In validation of the ventilator mode mean errors were -1.8% and -5.05% for VO2 and VCO2 with limits of agreement of 1.02 (+2SD) ml/min and -0.74 (-2SD) ml/min, and 0.93 (+2SD) ml/min and -1.45 (-2SD) ml/min respectively. Values of VO2 and VCO2 in 16 preterm infants in the canopy mode were 6.2 ml/kg/min (0.5 1SD) and 6.7 ml/kg/min (0.6 1SD) and in nine preterm infants in the ventilator mode 4.98 ml/kg/min (1.09 1SD) and 4.74 ml/min/kg (1.08 1SD) respectively. Mean energy expenditure was 45.5 kcal (191 kJ)kg/day for infants measured in the canopy mode and 35.5 kcal (149 kJ)/kg/day for ventilated infants. This metabolic system can be adapted for use in the newborn but accuracy is reduced when it is used in those weighing less than 1000 g.
机译:经验证,适合成人使用的便携式间接热量计可用于早产儿。随后在室内空气中自发呼吸的16名早产婴儿(冠层模式)和接受间歇性正压通气的9名早产婴儿(呼吸机模式)中测量了氧气消耗量(VO2)和二氧化碳生成量(VCO2)。该系统的验证是使用气体注入技术和氮气模拟VCO2的VO2和二氧化碳进行的。冠层模式验证的平均误差为VO2和VCO2为1.4%和0.2%,一致极限为0.6(+ 2SD)ml / min和-1.3(-2SD)ml / min,以及0.9(+ 2SD)ml / min分钟和-2.3(-2SD)毫升/分钟。在确认呼吸机模式时,VO2和VCO2的平均误差为-1.8%和-5.05%,一致极限为1.02(+ 2SD)ml / min和-0.74(-2SD)ml / min,以及0.93(+ 2SD) ml / min和-1.45(-2SD)ml / min。在冠层模式下的16例早产儿的VO2和VCO2值为6.2 ml / kg / min(0.5 1SD)和6.7 ml / kg / min(0.6 1SD),在9例在呼吸机模式下的早产儿4.98 ml / kg / min (1.09 1SD)和4.74 ml / min / kg(1.08 1SD)。在冠层模式下测得的婴儿平均能量消耗为45.5 kcal(191 kJ)/ kg /天,对于通气婴儿,平均能量消耗为35.5 kcal(149 kJ)/ kg / day。这种新陈代谢系统可以适合新生儿使用,但在体重不足1000克的人群中使用时,准确性会降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号