Improved thermometry system for ultrasound hyperthermia

机译:改进的超声热疗测温系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A thermometry system for use during ultrasound hyperthermia treatments was developed to provide fast and reliable temperature measurements such that transient temperatures from multi-sensor thermocouples could be measured. It was also intended to provide electrical isolation for patient safety when bare thermocouple sensors were used in order to reduce artifacts. The system hardware development involved fabrication of a high precision temperature measurement box which was electrically isolated from, by an opto-isolation unit, and interfaced with, an 386-20 MHz personal computer. The system software development involved a two point calibration program for each thermocouple probe to be used with the system, and a sensor locating program to rapidly identify the probe locations immediately prior to treatment. A single scan temperature reading speed of 0.2 sec for all 112 thermocouple sensors with an average accuracy of ±0.05°C under normal operating conditions (ambient temperature 22°C to 28°C) was achieved. A probe to earth ground leakage sink current of 75 μA and a leakage source current of less than 10 μA was attained.
机译:开发了一种在超声热疗期间使用的测温系统,以提供快速而可靠的温度测量,从而可以测量来自多传感器热电偶的瞬态温度。当使用裸露的热电偶传感器以减少伪影时,它还旨在为患者安全提供电气隔离。系统硬件开发涉及制造高精度温度测量箱,该测量箱通过光隔离单元与电隔离,并与386-20 MHz个人计算机接口。系统软件开发涉及针对要与系统一起使用的每个热电偶探针的两点校准程序,以及一个传感器定位程序,可以在治疗之前立即快速识别探针位置。在正常工作条件(环境温度22°C至28°C)下,所有112个热电偶传感器的单次扫描温度读取速度均为0.2秒,平均精度为±0.05°C。获得了一个对地泄漏漏电流为75μA且泄漏源电流小于10μA的探头。

著录项

  • 作者

    Lim Chuck Mang 1963-;

  • 作者单位
  • 年度 1992
  • 总页数
  • 原文格式 PDF
  • 正文语种 en_US
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号