Objective: To eliminate the effects of body deformation for MR-based prostate treatment planning, coil mounts are essential. In this study, we evaluated the effect of the coil set-up on image quality. Methods: A custom-designed pelvic-shaped phantom was scanned by systematically increasing the anterior body-to-coil (BTC) distance from 30 to 90 mm. The image quality near the organs of interest was determined in order to characterize the relationship between image quality and BTC distance at the critical organ structures. The half intensity reduction (HIR) was calculated to determine the sensitivity of each organ structure to the BTC distance change. Results: As the BTC distance increased, the uniformity reduced at 3% per millimetre. The HIR value indicated that the bladder signal is most sensitive to the change in BTC distance. By maintaining a constant BTC distance set-up, the intensity uniformity was improved by 28% along the B0 directions. Conclusion: Positioning the MRI coil on mounts can reduce body deformation but adversely degrades the image quality. The magnitude of this effect has been quantified for prostate MR simulation scanning. The coil needs to be positioned not only with a minimal but also uniform BTC distance in order to maximize image quality. Advances in knowledge: A method to characterize the effect on image quality due to the use of coil mounts was demonstrated. Coil mounts whose height can be adjusted individually to keep BTC distance constant are necessary to maintain a uniform image across the entire field of view.
机译:目的:为了消除身体变形对基于MR的前列腺治疗计划的影响,线圈架是必不可少的。在这项研究中,我们评估了线圈设置对图像质量的影响。方法:通过系统地将前体到线圈(BTC)的距离从30mm增加到90mm,扫描定制设计的骨盆形体模。确定感兴趣器官附近的图像质量,以表征关键器官结构处图像质量和BTC距离之间的关系。计算半强度降低(HIR),以确定每个器官结构对BTC距离变化的敏感性。结果:随着BTC距离的增加,均匀性降低了3%/毫米。 HIR值表明膀胱信号对BTC距离的变化最敏感。通过保持恒定的BTC距离设置,沿B0方向的强度均匀性提高了28%。结论:将MRI线圈放置在底座上可以减少身体变形,但会降低图像质量。已针对前列腺MR模拟扫描量化了这种影响的程度。为了最大化图像质量,不仅需要以最小而且要均匀的BTC距离定位线圈。知识的进步:展示了一种表征使用线圈架对图像质量的影响的方法。线圈架的高度可以单独调节以保持BTC距离恒定,这对于在整个视场中保持均匀的图像非常必要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号