首页> 外文OA文献 >Inter-rater reliability of automated devices for measurement of toe systolic blood pressure and the toe brachial index
【2h】

Inter-rater reliability of automated devices for measurement of toe systolic blood pressure and the toe brachial index

机译:自动测量脚趾收缩压和肱肱指数的评估者间可靠性

摘要

Objectives: The aim of this study was to determine the inter-rater reliability of automated devices for the measurement of toe systolic blood pressure (TSBP) and the toe brachial index (TBI). Participants and methods: Toe and brachial systolic blood pressures were measured in 40 participants by two podiatrists using SysToe and MicroLife automated devices. Intraclass correlation coefficients (ICC) with a 95% confidence interval (CI) and 95% limits of agreement (LOA) were determined for TSBP, brachial systolic blood pressure and the TBI. Results: The inter-rater reliability of all measurements was excellent, with ICCs ranging from 0.82 to 0.85. TSBP had the greatest reliability (ICC: 0.85, 95% CI: 0.74–0.92, 95% LOA: −14.4 to 21.1 mmHg) and brachial blood pressures had the lowest reliability (ICC: 0.82, 95% CI: 0.68–0.90, 95% LOA: −21.2 to 22.9 mmHg). The inter-rater reliability of the TBI fell between the TSBP and the brachial blood pressure reliability (ICC: 0.83, 95% CI: 0.70–0.90, 95% LOA: −0.16 to 0.20). Conclusion: We showed that automated devices for the measurement of toe and brachial systolic blood pressure and the TBI in a clinical setting produce measurements with high inter-rater reliability. These findings support the use of these automated devices for ongoing monitoring of lower extremity vascular status.
机译:目的:本研究的目的是确定用于测量脚趾收缩压(TSBP)和脚腕指数(TBI)的自动装置的评分间可靠性。参与者和方法:两名足病医生使用SysToe和MicroLife自动化设备在40名参与者中测量了脚趾和肱动脉收缩压。确定TSBP,肱动脉收缩压和TBI的类内相关系数(ICC),置信区间(CI)为95%和一致限(LOA)为95%。结果:所有测量的评估者间可靠性都非常好,ICC范围为0.82至0.85。 TSBP的可靠性最高(ICC:0.85,95%CI:0.74-0.92,95%LOA:-14.4至21.1mmHg),肱血压的可靠性最低(ICC:0.82,95%CI:0.68-0.90,95 LOA百分比:-21.2至22.9mmHg。 TBI的评分者间可靠性介于TSBP和肱血压可靠性之间(ICC:0.83,95%CI:0.70-0.90,95%LOA:-0.16至0.20)。结论:我们表明,在临床环境中,用于测量脚趾和肱动脉收缩压和TBI的自动化设备可产生具有较高评估者间可靠性的测量结果。这些发现支持将这些自动化设备用于持续监测下肢血管状态。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号