Calibration of photon and electron beams with an extrapolation chamber

机译:用外推室校准光子和电子束

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A variable air-volume, parallel-plate, extrapolation chamber forming an integral part of a Solid-WaterTM phantom was built to determine the absorbed dose in Solid-WaterTM directly. The sensitive air-volume of the extrapolation chamber is controlled through the movement of the chamber piston by means of a micrometer mounted to the phantom body. The relative displacement of the piston is monitored by a mechanical distance travel indicator with a precision on the order of 0.002 mm. Irradiations were carried out with cobalt-60 gamma. rays, x-ray beams ranging from 4 to 18 MV, and electron beams between 6 and 22 MeV. The absorbed dose at a given depth in Solid-WaterTM is proportional to the ionization gradient measured in the Bragg-Gray cavity region with a Solid-WaterTM embedded extrapolation chamber. Measured charge is corrected for ion recombination and ion diffusion in the chamber air volume according to a comprehensive model for charge loss in an ionization chamber. The discrepancies between doses determined with our uncalibrated phantom-embedded extrapolation chamber and doses obtained with calibrated Farmer-type cylindrical ionization chambers following the AAPM-TG21 and AAPM-TG25 dosimetry protocols are less than 1% for photon and electron beams at all clinical megavoltage energies. Uncalibrated extrapolation chambers thus offer a simple and practical alternative to other techniques used in output measurements of megavoltage photon and electron beams.
机译:建立了可变风量,平行板外推腔,该腔构成Solid-WaterTM幻像的组成部分,可直接确定Solid-WaterTM的吸收剂量。通过安装在体模上的千分尺通过室活塞的运动来控制外推室的灵敏空气量。活塞的相对位移由机械距离行程指示器监控,精度为0.002 mm。用钴60γ射线进行辐照。射线,4到18 MV的X射线束和6到22 MeV的电子束。在Solid-WaterTM中给定深度处的吸收剂量与在带有Solid-WaterTM嵌入式外推腔的布拉格灰色腔区域中测得的电离梯度成比例。根据用于电离室中电荷损失的综合模型,对测得的电荷进行校正,以便在室空气量中进行离子重组和离子扩散。对于所有临床兆伏能量下的光子和电子束,使用AAPM-TG​​21和AAPM-TG​​25剂量学规程的未经校准的幻像嵌入式外推室确定的剂量与经校准的Farmer型圆柱电离室得到的剂量之间的差异小于1% 。因此,未经校准的外推腔可为兆伏光子和电子束输出测量中使用的其他技术提供简单实用的替代方案。

著录项

  • 作者

    Zankowski, Corey E.;

  • 作者单位
  • 年度 1997
  • 总页数
  • 原文格式 PDF
  • 正文语种 en
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号