Fetal head position and perineal distension associated with the use of the BD Odon DeviceTM in operative vaginal birth:a simulation study

机译:BD Odon Device TM 在手术中阴道分娩中的胎儿头部位置和会阴扩张的模拟研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

ObjectiveTo investigate (i) the placement of the BD Odon Device on the model fetal head, and (ii) perineal distention during simulated operative vaginal births conducted with the BD Odon DeviceDesignObservational simulation studySettingNorth Bristol NHS Trust, UKPopulation or Sample440 simulated operative vaginal birthsMethodsThree bespoke fetal mannequins were developed to represent (i) bi-parietal diameter of the 50th centile at term (ii) bi-parietal diameter at the 5th centile at term and (iii) 50th centile head with 2 cm of caput. Siting of the BD Odon Device on model heads was determined before and after 400 simulated operative vaginal births. Variables were analysed to determine their effect on device siting and movement during birth. The fetal mannequins were placed inside a maternal mannequin (PROMPT Flex, Limbs & Things, Bristol, UK) and the BD Odon Device was placed around the fetal head as per the instructions for use. The location of the air cuff was determined before and after the head was delivered. Perineal distension was determined by recording maximum perineal distention during a simulated operative vaginal birth using the same procedure, as well as scenarios employing an inappropriately non-deflated air cuff (for the BD Odon Device), the Kiwi ventouse and non-rotational forceps.Main Outcome Measures Site and displacement during birth of the BD Odon Device on a model head. Maximal perineal distension during birth.Results The BD Odon Device was reliably sited in a standard over the fetal head position (approximately 40mm above the fetal chin) for all stations, head sizes and positions with no significant displacement. In occipito-posterior births, compared to occipito-anterior or transverse, the BD Odon Device routinely sited further down the fetal head (toward the chin).The BD Odon Device was not associated with more perineal distension than forceps or Kiwi ventouse (21mm vs 26mm vs 21mm at posterior fourchette). ConclusionsThe BD Odon Device reliably sited over a safe area of the fetal head in 400 simulated births representative of clinical practice. The BD Odon Device generates similar levels of perineal distension compared to Kiwi ventouse when used correctly
机译:目的研究(i)将BD Odon Device放置在模型胎头上,以及(ii)使用BD Odon DeviceDesign进行模拟手术阴道分娩时的会阴部扩张观察模拟研究设置North Bristol NHS Trust,UKPopulation或Sample440模拟手术阴道分娩的方法三定制胎儿人体模型被开发为代表(i)足月第50个百分位的顶壁直径(ii)足月第5个百分位的顶壁直径和(iii)头高2 cm的第50个百分头。在模拟的400例手术阴道分娩前后,确定BD Odon设备在模型头上的位置。分析变量以确定其对出生时设备选址和移动的影响。将胎儿人体模型放置在产妇人体模型(PROMPT Flex,Limbs&Things,英国布里斯托尔)内,并按照使用说明将BD Odon Device放置在胎儿头部周围。在输送头之前和之后确定气囊的位置。通过记录在模拟手术阴道分娩过程中最大的会阴部扩张来确定会阴部扩张,使用相同的步骤,以及使用不适当的不放气的袖带(用于BD Odon设备),猕猴桃通气管和不旋转的镊子的情况。结果测量BD Odon设备在模型头上出生时的部位和移位。出生时会阴部最大扩张。结果BD Odon设备可靠地放置在胎儿头位置(胎儿下巴上方约40mm)的标准位置,所有位置,头部大小和位置均无明显移位。与枕后或横枕相比,枕后分娩中的BD Odon装置通常位于胎儿头部下方(朝下巴)。与镊子或猕猴桃腹腔相比,BD Odon装置与会阴扩张的发生率无关(21mm vs.后方切线:26mm和21mm)。结论BD Odon设备可靠地放置在代表临床实践的400例模拟出生的胎儿头部安全区域内。如果正确使用,BD Odon设备会产生与猕猴桃腹膜相似的会阴膨胀程度

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号