A dynamic balloon volume sensor system (DBVSS) was designed for use with the intra-aortic balloon pump (IABP), a therapeutic device to assist heart recovery after cardiac dysfunction or cardiac trauma, and the Pittsburgh respiratory support catheter (RSC), an internally deployed gas exchange device which augments lung function. The DBVSS was designed to detect the degree of inflation of the balloons incorporated into each device as they pulse within a patient. Both devices require full inflation for optimal performance, and both will under-inflate during normal operation. The sensor system requirements were to measure volumes within 10% of the actual across the range of expected pulsation frequencies as well as in changing concentrations of helium.The DBVSS employed a hot wire anemometer to detect the flow entering the balloon, combined with a computer algorithm to integrate the flow to find volume. The system compensated for the flow reading changes resulting from changing helium concentration by measuring gas properties during zero gas flow between pulsations, and used this data to correct the flow profile at each helium concentration. The volume from the DBVSS was compared to the volume standard as measured by water displacement in a plethysmograph.The system was able to accurately measure delivered balloon volume under changing gas composition as well as detected volume loss from the balloon across helium concentrations. The DBVSS measured the volume within 10% across these tests, as well as under compression of the balloon, high resistance in the driveline and across frequencies up to 480 beats per minute. The DBVSS was proved to be within the design requirements for helium concentration and inflation methods for both the devices considered.
机译:动态球囊容积传感器系统(DBVSS)设计用于与主动脉内球囊泵(IABP)(一种在心脏功能障碍或心脏创伤后协助心脏恢复的治疗设备)以及匹兹堡呼吸支持导管(RSC)一起使用部署了可以增强肺功能的气体交换设备。 DBVSS旨在检测在患者体内脉动时合并到每个设备中的球囊的充气程度。两种设备都需要完全充​​气才能获得最佳性能,并且两种设备在正常运行期间都会充气不足。传感器系统的要求是在预期的脉动频率范围内以及在不断变化的氦气浓度下测量体积在实际体积的10%以内.DBVSS使用热线风速计检测进入气球的流量,并结合计算机算法整合流量以找到体积。该系统通过在脉动之间零气体流量期间测量气体特性来补偿由于氦浓度变化而引起的流量读数变化,并使用该数据校正每个氦浓度下的流量分布。将DBVSS的体积与体积描记器中通过水置换法测量的体积标准进行比较。该系统能够准确测量在不断变化的气体成分下输送的球囊体积,并检测跨氦气浓度的球囊体积损失。在这些测试中,以及在气球压缩,动力传动系统中的高阻力以及高达每分钟480次搏动的频率下,DBVSS都测量了10%以内的音量。事实证明,DBVSS符合所考虑的两种器件的氦气浓度和充气方法的设计要求。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号