Alterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate

机译:手动操作的自充气式复苏袋所产生的吸气峰值压力和潮气量随供氧量的变化而变化

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVE: To assess possible alterations in the tidal volume and peak inspiratory pressure delivered by seven models of manually operated self-inflating resuscitation bags as a function of the oxygen supply rate used. METHODS: The resuscitation bags tested were the following: Oxigel, models A and B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; and Narcosul. For the measurements, a wall oxygen flow meter, a flow meter/respirometer, a resuscitation bag, a sensor (Tracer 5 unit), and a test lung were connected. In addition, the Tracer 5 unit was connected to a notebook computer. Oxygen supply rates of 1, 5, 10, and 15 L/min were used. RESULTS: The tidal volume delivered by the Oxigel model A resuscitation bag when receiving oxygen at a rate of 15 L/min was approximately 99% greater than that delivered when receiving oxygen at a rate of 1 L/min. Similarly, peak inspiratory pressure was approximately 155% greater. Under the same conditions, the tidal volume delivered by the Narcosul resuscitation bag was 48% greater, and peak inspiratory pressure was 105% greater. The remaining resuscitation bags tested showed no significant alterations in the tidal volume or peak inspiratory pressure delivered. CONCLUSIONS: Under the resistance and compliance conditions used, the resuscitation bags in which the oxygen inflow is directly to the interior of the bag had the patient valve stuck at the inspiratory position when receiving oxygen at a rate > 5 L/min, significantly increasing the tidal volume and peak inspiratory pressure delivered. This did not occur with the resuscitation bags in which the oxygen inflow is directed to the exterior of the bag.
机译:目的:评估七个模型的手动充气式自充气复苏袋所提供的潮气量和峰值吸气压力随所用氧气供应率的变化。方法:测试的复苏袋如下:Oxigel,型号A和B; CE Reanimadores; ProtecSolutions;密苏里州斧头和Narcosul。为了进行测量,连接了壁式氧气流量计,流量计/呼吸计,复苏袋,传感器(Tracer 5单元)和肺部测试。此外,Tracer 5单元已连接到笔记本计算机。供氧速率为1、5、10和15 L / min。结果:Oxigel A型复苏袋在以15 L / min的速度吸入氧气时的潮气量比以1 L / min的速度在氧气吸收时潮气量大约99%。同样,峰值吸气压力大约增加了155%。在相同条件下,Narcosul复苏袋输送的潮气量增加了48%,峰值吸气压力增加了105%。其余测试的复苏袋显示潮气量或峰值吸气压力无明显变化。结论:在所用的阻力和顺应性条件下,当氧气以> 5 L / min的速度接收时,氧气直接流入气囊内部的复苏袋将患者瓣膜卡在吸气位置,潮气量和峰值吸气压力。对于复苏袋,氧气流入袋的外部不会发生这种情况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号