首页> 外文OA文献 >Pulsatile Stent Graft: A New Alternative In Chronic Ventricular Assistance [stent Aórtico Pulsátil: Uma Nova Alternativa Na Assistência Ventricular Crônica]
【2h】

Pulsatile Stent Graft: A New Alternative In Chronic Ventricular Assistance [stent Aórtico Pulsátil: Uma Nova Alternativa Na Assistência Ventricular Crônica]

机译:搏动性支架移植:慢性心室协助的新选择[搏动性主动脉支架:慢性心室协助的新选择]

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective: Heart failure is currently one of the most common hospitalization causes. Several chronic circulatory assist devices have been tested and are highly complex. The objective is the description of a pulsatile endoprosthesis capable of applying a chronic pulse within the descending aorta, similar to that produced by intra-aortic balloon. Methods: Pulsatile stents composed of nickel-titanium were built and positioned to engage latex tubes simulating the aorta. Different electric currents were applied to units connected in series in order to cause structure contraction and displacement of a liquid column. There were two sequence tests: first composed of two metallic cages and the second composed of five cages. At first sequence tests was applied a voltage of 16.3 volts and a current of 5 amperes. In the second, voltage of 15 volts and current of 07 amperes. Results: In the first sequence was obtained the pulsatile effect of stent, with contraction of the tube and displacement of the water column sufficient to validate the pulsating effect of the endoprosthesis. The two structures ejected a volume of 2.6 ml per cycle, with a range of 29 mm in height of the column of water equivalent to 8% shrinkage during the pulse. In the second sequence, it reachead a variation of 7.4 mL per cycle. Conclusion: The results obtained confirm the stent pulsatile contractility activated by electrical current. The continuity of the study and material improvement are necessary to obtain more efficient model from the point of view of energy and pulse, to allow ejection volumes comparable with the intra-aortic balloons.
机译:目的:心力衰竭是目前最常见的住院原因之一。已经测试了几种慢性循环辅助设备,它们非常复杂。目的是描述一种能够在降主动脉内施加慢性脉搏的搏动性内置假体,类似于由主动脉内球囊产生的那种。方法:构建由镍钛合金制成的脉冲支架,并放置在与模拟主动脉的乳胶管接合的位置。将不同的电流施加到串联连接的单元上,以引起结构收缩和液柱移位。有两个顺序测试:第一个由两个金属笼组成,第二个由五个笼组成。在第一个序列测试中,施加了16.3伏的电压和5安培的电流。在第二个中,电压为15伏,电流为07安培。结果:在第一个序列中,获得了支架的搏动效果,其中管的收缩和水柱的位移足以验证内假体的搏动效果。这两个结构每个循环喷射的体积为2.6 ml,水柱的高度范围为29 mm,相当于在脉冲过程中收缩了8%。在第二个序列中,每个循环的变化幅度为7.4 mL。结论:获得的结果证实了电流激活的支架搏动收缩性。从能量和脉冲的角度来看,研究的连续性和材料的改进对于获得更有效的模型是必要的,以允许与主动脉内球囊相当的射血量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号