Accuracy and repeatability of quantitative fluoroscopy for the measurement of sagittal plane translation and finite centre of rotation in the lumbar spine

机译:定量荧光透视术在腰椎矢状面平移和有限旋转中心测量中的准确性和可重复性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Quantitative fluoroscopy (QF) was developed to measure intervertebral mechanics in vivo and has been found to have high repeatability and accuracy for the measurement of intervertebral rotations. However, sagittal plane translation and finite centre of rotation (FCR) are potential measures of stability but have not yet been fully validated for current QF. This study investigated the repeatability and accuracy of QF for measuring these variables. Repeatability was assessed from L2-S1 in 20 human volunteers. Accuracy was investigated using 10 consecutive measurements from each of two pairs of linked and instrumented dry human vertebrae as reference; one which tilted without translation and one which translated without tilt. The results found intra- and inter-observer repeatability for translation to be 1.1mm or less (SEM) with fair to substantial reliability (ICC 0.533-0.998). Intra-observer repeatability of FCR location for inter-vertebral rotations of 5o and above ranged from 1.5mm to 1.8mm (SEM) with moderate to substantial reliability (ICC 0.626-0.988). Inter-observer repeatability for FCR ranged from 1.2mm to 5.7mm, also with moderate to substantial reliability (ICC 0.621-0.878). Reliability was substantial (ICC>0.81) for 10/16 measures for translation and 5/8 for FCR location. Accuracy for translation was 0.1mm (fixed centre) and 2.2mm (moveable centre), with an FCR error of 0.3mm(x) and 0.4mm(y) (fixed centre). This technology was found to have a high level of accuracy and with a few exceptions, moderate to substantial repeatability for the measurement of translation and FCR from fluoroscopic motion sequences.
机译:定量荧光检查法(QF)的开发是为了测量体内椎体力学,并且发现其具有很高的可重复性和准确性,可用于测量椎间旋转。但是,矢状面平移和有限旋转中心(FCR)是潜在的稳定性度量,但尚未针对当前QF进行充分验证。这项研究调查了测量这些变量的QF的可重复性和准确性。从L2-S1评估了20位人类志愿者的可重复性。使用两对相连的和人工干燥的人类椎骨对中的每对进行10次连续测量,以调查准确性。一种不倾斜而倾斜,另一种不倾斜而平移。结果发现,观察者之间和观察者之间的翻译重复性为1.1mm或更小(SEM),相当可靠(ICC 0.533-0.998)。椎间旋转5o及以上时,FCR位置的观察者可重复性在1.5mm至1.8mm(SEM)范围内,具有中等到相当高的可靠性(ICC 0.626-0.988)。 FCR的观察者间重复性在1.2mm至5.7mm的范围内,并且具有中等到充分的可靠性(ICC 0.621-0.878)。对于10/16的翻译量度和对FCR定位的5/8量度,可靠性是很高的(ICC> 0.81)。平移精度为0.1mm(固定中心)和2.2mm(可移动中心),FCR误差为0.3mm(x)和0.4mm(y)(固定中心)。发现该技术具有很高的准确性,除了少数例外,对于荧光透视运动序列的翻译和FCR的测量具有中等至相当的可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号