Inter- and intra-observer variability analysis of completely automated cIMT measurement software (AtheroEdge™) and its benchmarking against commercial ultrasound scanner and expert Readers

机译:全自动cIMT测量软件(AtheroEdge™)的观察者之间和观察者内部变异性分析以及其对商业超声扫描仪和专家阅读器的基准测试

摘要

The purpose of this study was to evaluate the measurement error and inter- and intra-observer variability of completely off-line automated and semi-automated carotid intima-media thickness (cIMT) measurement software (AtheroEdge™). Two hundred carotid ultrasound images from 50 asymptomatic women were analyzed. AtheroEdge™ was benchmarked against a commercial system (Syngo, Siemens) using automated and semi-automated modes. The measurement error and inter- and intra-observer variability of AtheroEdge™ were tested using three readings. The measurement error of AtheroEdge™ compared to the commercial software was 0.002±0.019mm (r=0.99) in the automated mode and -0.001±0.004mm in the semi-automated mode (r=0.99). The measurement error of AtheroEdge™ compared to the mean value of the three expert Readers (cIMT bias) for the automated and semi-automated methods was -0.0004±0.158mm and -0.008±0.157mm, respectively. The Figure-of-Merit was 99.8% and 99.9% when compared to the commercial ultrasound scanner (using the automated and semi-automated method, respectively) and was 99.9% and 98.9% when compared to the mean value of the three expert Readers. Regarding inter- and intra-observer variability, the intra-class correlation coefficient of the three independent users using the semi-automated AtheroEdge™ was 0.98. AtheroEdge™ showed a measurement performance comparable to the commercial ultrasound scanner software and the expert Readers' tracings. AtheroEdge™ belongs to a class of automated systems that could find application in processing large datasets for common carotid arteries, avoiding subjectivity in cIMT measurements.
机译:这项研究的目的是评估完全离线的自动和半自动颈动脉内膜中膜厚度(cIMT)测量软件(AtheroEdge™)的测量误差以及观察者之间和观察者内部的变异性。分析了来自50名无症状女性的200例颈动脉超声图像。 AtheroEdge™在使用自动和半自动模式的商业系统(Syngo,Siemens)上进行了基准测试。使用三个读数测试了AtheroEdge™的测量误差以及观察者之间和观察者内部的变异性。与商业软件相比,AtheroEdge™的测量误差在自动模式下为0.002±0.019mm(r = 0.99),在半自动模式下为-0.001±0.004mm(r = 0.99)。与自动和半自动方法的三个专家读数器的平均值(cIMT偏差)相比,AtheroEdge™的测量误差分别为-0.0004±0.158mm和-0.008±0.157mm。与商用超声扫描仪(分别使用自动和半自动方法)相比,品质因数为99.8%和99.9%,与三个专家阅读器的平均值相比,分别为99.9%和98.9%。关于观察者之间和观察者内部的变异性,使用半自动化的AtheroEdge™的三个独立用户的类内相关系数是0.98。 AtheroEdge™的测量性能可与商用超声扫描仪软件和专业的Readers追踪相媲美。 AtheroEdge™属于一类自动化系统,可以在处理大型颈总动脉数据集时找到应用,从而避免了cIMT测量的主观性。

著录项

相似文献

  • 外文文献
  • 中文文献

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号