A sensitive approach, based on semi-quantitative measurement of the characteristic fragments in multistage extractive electrospray ionization mass spectrometry (EESI-MSn), was developed for fast detection of trace levels of lead in aqueous liquids including mineral water, lake water, tap water, energy drinks, soft drinks, beer, orange juice, and tea. A disodium ethylene-diamine-tetraacetic acid (EDTA) aqueous solution was electrosprayed to produce negatively charged primary ions which then intersected the neutral sample plume to generate anions of EDTA-Pb(II) complexes. The charged EDTA-Pb(II) complexes were characterized with multistage collision induced dissociation (CID) experiments. The limit of detection (LOD) using EESI-MS3 was estimated to be at the level of 10(-13) g/mL for directly detecting lead in many of these samples. The linear dynamic range was higher than 2 orders of magnitude. A single sample analysis could be completed within 2 min with reasonable semi-quantitative performance, e.g., relative standard deviations (RSDs) for deionized water were 4.6-7.6% during 5 experimental runs (each of them had 10 repeated measurements). Coca-cola and Huiyuan orange juice, representative beverage samples with complex matrices, generated recovery rates of 91.5% and 129%, respectively. Our experimental data demonstrated that EESI-MS is a useful tool for the fast detection of lead in various solutions, and EESI-MS showed promises for fast screening of lead-contaminated aqueous liquid samples. (C) 2012 Elsevier B.V. All rights reserved,
机译:开发了一种灵敏的方法,该方法基于多级萃取电喷雾电离质谱(EESI-MSn)中特征片段的半定量测量,可用于快速检测矿泉水,湖水,自来水,能量饮料,汽水,啤酒,橙汁和茶。乙二胺二胺四乙酸二钠(EDTA)水溶液被电喷雾产生带负电荷的初级离子,然后与中性样品羽流相交以生成EDTA-Pb(II)络合物的阴离子。带电的EDTA-Pb(II)配合物通过多阶段碰撞诱导解离(CID)实验进行了表征。为了直接检测其中许多样品中的铅,估计使用EESI-MS3的检出限(LOD)为10(-13)g / mL。线性动态范围高于2个数量级。可以在2分钟内以合理的半定量性能完成一次样品分析,例如,在5个实验运行中,去离子水的相对标准偏差(RSD)为4.6-7.6%(每个重复测量10次)。具有复杂基质的代表性饮料样品可口可乐和汇源橙汁的回收率分别为91.5%和129%。我们的实验数据表明,EESI-MS是用于快速检测各种溶液中铅的有用工具,并且EESI-MS显示了对铅污染的水性液体样品进行快速筛查的希望。 (C)2012 Elsevier B.V.保留所有权利,

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号