Development of gas chromatography-flame ionization detection system with a single column and liquid nitrogen-free for measuring atmospheric C2-C12 hydrocarbons

机译:单柱无液氮气相色谱-火焰电离检测系统的开发,用于测量大气中的C2-C12碳氢化合物

摘要

A liquid nitrogen-free GC-FID system equipped with a single column has been developed for measuring atmospheric C2-C12 hydrocarbons. The system is consisted of a cooling unit, a sampling unit and a separation unit. The cooling unit is used to meet the temperature needs of the sampling unit and the separation unit. The sampling unit includes a dehydration tube and an enrichment tube. No breakthrough of the hydrocarbons was detected when the temperature of the enrichment tube was kept at -90 degrees C and sampling volume was 400 mL. The separation unit is a small round oven attached on the cooling column. A single capillary column (0V-1,30 m x 0.32 mm I.D.) was used to separate the hydrocarbons. An optimal program temperature (-60 similar to 170 degrees C) of the oven was achieved to efficiently separate C2-C12 hydrocarbons. There were good linear correlations (R-2=0.993-0.999) between the signals of the hydrocarbons and the enrichment amount of hydrocarbons, and the relative standard deviation (RSD) was less than 5%, and the method detection limits (MDLs) for the hydrocarbons were in the range of 0.02-0.10 ppbv for sampling volume of 400 mL. Field measurements were also conducted and more than 50 hydrocarbons from C2 to C12 were detected in Beijing city. (C) 2015 Elsevier B.V. All rights reserved.
机译:已开发出一种配备单柱的无液氮GC-FID系统,用于测量大气中的C2-C12碳氢化合物。该系统由冷却单元,采样单元和分离单元组成。冷却单元用于满足采样单元和分离单元的温度需求。采样单元包括脱水管和浓缩管。当浓缩管的温度保持在-90摄氏度且采样量为400毫升时,未检测到碳氢化合物的穿透。分离单元是安装在冷却塔上的小型圆形烤箱。使用单个毛细管柱(0V-1,30 m x 0.32 mm I.D.)分离烃。实现了烤箱的最佳程序温度(-60类似于170摄氏度)以有效分离C2-C12碳氢化合物。碳氢化合物的信号与碳氢化合物的富集量之间具有良好的线性相关性(R-2 = 0.993-0.999),相对标准偏差(RSD)小于5%,并且方法检测限(MDLs)对于400 mL的采样量,碳氢化合物的浓度范围为0.02-0.10 ppbv。还进行了现场测量,并在北京市检测到了从C2到C12的50多种碳氢化合物。 (C)2015 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号