Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).

机译:使用Vero4DRT(MHI-TM2000)的万向接头x射线头对基于红外标记的动态肿瘤跟踪辐射进行准确性验证。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

[Purpose] To verify the accuracy of an infrared (IR) marker-based dynamic tumor-tracking irradiation system (IR tracking) using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). Conclusions: Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error [Conclusions] Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error.
机译:[目的]使用Vero4DRT(MHI-TM2000)的万向X射线头验证基于红外(IR)标记的动态肿瘤跟踪辐射系统(IR跟踪)的准确性。结论:红外跟踪中的位置跟踪误差与4D建模误差密切相关。因此,必须在处理之前验证4D模型的准确性,并且需要余量来补偿4D建模误差。[结论]位置跟踪误差与IR跟踪中的4D建模误差密切相关。因此,必须在处理之前验证4D模型的准确性,并且需要余量来补偿4D建模误差。