首页> 外文OA文献 >Precision closed bomb calorimeter for testing flame and gas producing initiators
【2h】

Precision closed bomb calorimeter for testing flame and gas producing initiators

机译:精密封闭式炸弹量热仪,用于测试产生火焰和气体的引发剂

摘要

A calorimeter has been developed under this study to help meet the needs of accurate performance monitoring of electrically or mechanically actuated flame and gas producing devices, such as squib-type initiators. A ten cubic centimeter closed bomb (closed volume) calorimeter was designed to provide a standard pressure trace and to measure a nominal 50 calorie output, using the basic components of a Parr Model 1411 calorimeter. Two prototype bombs were fabricated, pressure tested to 2600 psi, and extensively evaluated.
机译:根据这项研究,已开发出一种量热仪,可满足对电动或机械驱动的火焰和气体产生装置(例如爆管型起爆器)进行精确性能监控的需求。设计了一个十立方厘米的封闭式炸弹(封闭体积)量热计,以使用Parr 1411型量热计的基本组件来提供标准压力迹线并测量名义上的50卡热量输出。制造了两枚原型炸弹,压力测试至2600 psi,并进行了广泛的评估。

著录项