A combination of a single mode AlGaAs laser diode and broadband LED was used in a Michelson interferometer to provide reference signals in a Fourier transform spectrometer, the Composite Infrared Spectrometer, on the Cassini mission to Saturn. The narrowband light from the laser produced continuous fringes throughout the travel of the interferometer, which were used to control the velocity of the scan mechanism and to trigger data sampling. The broadband light from the LED produced a burst of fringes at zero path difference, which was used as a fixed position reference. The system, including the sources, the interferometer, and the detectors, was designed to work both at room temperature and instrument operating temperature of 170 Kelvin. One major challenge that was overcome was preservation, from room temperature to 170 K, of alignment sufficient for high modulation of fringes from the broadband source. Another was the shift of the source spectra about 30 nm toward shorter wavelengths upon cooldown.
机译:在迈克尔逊干涉仪中,将单模AlGaAs激光二极管和宽带LED的组合用于傅里叶变换光谱仪(复合红外光谱仪)中的参考信号,以执行卡西尼号对土星的任务。来自激光的窄带光在干涉仪的整个行程中产生连续的条纹,这些条纹用于控制扫描机制的速度并触发数据采样。来自LED的宽带光在零光程差处产生了一阵条纹,该条纹被用作固定位置参考。该系统包括信号源,干涉仪和检测器,旨在在室温和170开尔文的仪器工作温度下工作。克服的一个主要挑战是在室温至170 K的温度范围内保持足够的对准度,以便对宽带源的条纹进​​行高调制。另一个是冷却后,光源光谱向短波长方向移动了约30 nm。