Using a NPWE model observer to assess suitable image quality for a digital mammography quality assurance programme

机译:使用NPWE模型观察器评估数字乳腺摄影质量保证程序的合适图像质量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A method of objectively determining imaging performance for a mammography quality assurance programme for digital systems was developed. The method is based on the assessment of the visibility of a spherical microcalcification of 0.2 mm using a quasi-ideal observer model. It requires the assessment of the spatial resolution (modulation transfer function) and the noise power spectra of the systems. The contrast is measured using a 0.2-mm thick Al sheet and Polymethylmethacrylate (PMMA) blocks. The minimal image quality was defined as that giving a target contrast-to-noise ratio (CNR) of 5.4. Several evaluations of this objective method for evaluating image quality in mammography quality assurance programmes have been considered on computed radiography (CR) and digital radiography (DR) mammography systems. The measurement gives a threshold CNR necessary to reach the minimum standard image quality required with regards to the visibility of a 0.2-mm microcalcification. This method may replace the CDMAM image evaluation and simplify the threshold contrast visibility test used in mammography quality
机译:开发了一种客观确定数字系统乳腺摄影质量保证程序成像性能的方法。该方法基于使用准理想观察者模型评估的0.2毫米球形微钙化的可见性。它需要评估系统的空间分辨率(调制传递函数)和噪声功率谱。使用0.2毫米厚的Al板和聚甲基丙烯酸甲酯(PMMA)块测量对比度。最低图像质量定义为目标对比度/噪声比(CNR)为5.4。在计算机X线摄影(CR)和数字X线摄影(DR)乳腺摄影系统上,已经考虑了该客观方法在乳腺摄影质量保证程序中评估图像质量的几种评估。该测量给出了就0.2毫米微钙化的可见度而言达到所需的最低标准图像质量所需的阈值CNR。此方法可以代替CDMAM图像评估,并简化用于乳房X线照相质量的阈值对比度可见性测试

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号