Abstract Objective: To determine the in vitro efficacy of glucoprotamin for the disinfection of instruments. Design: Prospective observational study. Setting: University women's hospital. Methods: Instruments were immersed in saline solution after use, and glucoprotamin was added to a concentration of 1.5% before soaking for 60 minutes. Biocidal activity was determined by the difference in colony-forming units (CFU) on instruments before and after disinfection. Results: One hundred thirty-seven instruments were collected during 10 days and exposed to a 1.5% dilution of glucoprotamin without prior washing. Bioburden before disinfection ranged from 2 × 105 to 7.1 × 107 CFU per instrument. Average bacterial killing was 5.98 log10 CFU ± 0.48 under aerobic conditions and 6.75 log10 CFU ± 0.54 under anaerobic conditions, despite the presence of large amounts of proteins on instruments that were frequently bloody. No vegetative bacteria were isolated in any sample after disinfection. Conclusion: This clinical study confirmed excellent in vitro efficacy of glucoprotamin without prior removal of proteins and debris
机译:摘要目的:确定葡萄糖原蛋白在器械消毒中的体外功效。设计:前瞻性观察研究。地点:大学女子医院。方法:使用后将器械浸入盐水溶液中,并在浸泡60分钟之前加入1.5%的葡萄糖原蛋白。杀菌活性由消毒前后仪器上菌落形成单位(CFU)的差异确定。结果:在10天内收集了一百三十七种仪器,并且未经事先清洗就暴露于1.5%的葡萄糖原蛋白稀释液中。每台仪器在消毒前的生物负荷范围为2×105至7.1×107 CFU。尽管在经常流血的器械上存在大量蛋白质,但在有氧条件下,平均细菌杀死率为5.98 log10 CFU±0.48,在无氧条件下为6.75 log10 CFU±0.54。消毒后在任何样品中都没有分离出营养细菌。结论:该临床研究证实了葡糖蛋白的优异体外功效,而无需事先去除蛋白质和碎屑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号