首页> 外文OA文献 >Development of the ’BreathEasy’ contactless portable respiratory rate monitor (CPRM)
【2h】

Development of the ’BreathEasy’ contactless portable respiratory rate monitor (CPRM)

机译:开发“ BreathEasy”非接触式便携式呼吸速率监测器(CPRM)

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: Respiratory rate (RR) is a vital physiological measurement used in the immediate assessment of acutely ill patients. It is used as a predictor of serious deterioration in a patient's clinical condition. Convenient electronic devices exist for measurement of pulse, blood pressure, oxygen saturation and temperature. Although devices which measure RR exist, none has entered everyday clinical practice. We have developed a contactless portable respiratory rate monitor (CPRM). We aimed to measure agreement with existing methods of RR measurement.Method:RR data were collected from 33 adult volunteers using respiratory impedance plethysmography (RIP) bands (established contact method), visual counting of chest movements (established non-contact method) and the CPRM (new method), simultaneously. Two to three data sets were collected for each volunteer.Results:Data showed good agreement between measurements from the CPRM and the gold standard RIP with limits of agreement -4.6 – 7.8.
机译:背景:呼吸频率(RR)是用于立即评估急性病患者的重要生理指标。它被用作预测患者临床状况严重恶化的指标。存在方便的电子设备,用于测量脉搏,血压,血氧饱和度和温度。尽管存在可测量RR的设备,但尚未进入日常临床实践。我们已经开发了一种非接触式便携式呼吸频率监测器(CPRM)。我们的目的是与现有的RR测量方法进行一致性测量。方法:使用呼吸阻抗体积描记术(RIP)频段(已建立的接触方法),肉眼观察胸部运动(已建立的非接触方法)和33例成人志愿者收集RR数据。同时使用CPRM(新方法)。为每位志愿者收集了两到三个数据集。结果:数据显示CPRM和金标准RIP的测量值之间具有良好的一致性,一致性极限为-4.6 – 7.8。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号