【2h】

Performance evaluation of the FLEX triumph X-PET scanner using the national electrical manufacturers association NU-4 standards

机译:使用国家电气制造商协会NU-4标准的FLEX胜利X-PET扫描仪的性能评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this work was to evaluate the performance characteristics of the preclinical X-PET subsystem of the FLEX Triumph PET/CT scanner based on the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). METHODS: The performance parameters evaluated include the spatial resolution, scatter fraction, count losses and random coincidences, sensitivity, and image-quality characteristics. The PET detector array consisted of 11,520 individual bismuth germanate crystals arranged in 48 rings and 180 blocks, with an axial field of view (FOV) of 11.6 cm and a inner ring diameter of 16.5 cm. The spatial resolution was measured with a small (22)Na point source (diameter, 0.25 mm) at different radial offsets from the center. Sensitivity was calculated using the same source by stepping the source axially through the axial FOV of the scanner. Scatter fraction and counting-rate performances were determined using a mouse- and rat-sized phantom with an (18)F line source insert. The NEMA image-quality phantom and rodent imaging were also performed to access the overall imaging capabilities of the scanner. RESULTS: Tangential spatial resolution in terms of full width at half maximum varied between 2.2 mm at the center of the FOV and 2.3 mm at a radial offset of 2.5 cm. The radial spatial resolution varied between 2.0 at the center and 4.4 mm at a radial offset of 2.3 cm. The peak system absolute sensitivity was 5.9% at the center of the FOV. The absolute system sensitivity was 0.67 counts/s/Bq, and the relative total system sensitivity was 73.9%. The scatter fraction for the mouse-sized phantom was 7.9%, with a peak true counting rate of 168 kilocounts per second (kcps) at 0.3 MBq/mL and a peak noise-equivalent counting rate of 106 kcps at 0.17 MBq/mL. The rat-sized phantom had a scatter fraction of 21%, with a peak true counting rate of 93 kcps at 0.034 MBq/mL and a peak noise-equivalent counting rate of 49 kcps at 0.02 MBq/mL. Recovery coefficients for the image-quality phantom ranged from 0.13 to 0.88. CONCLUSION: The performance of the X-PET scanner based on the NEMA NU 4-2008 standards was fully characterized. The overall performance demonstrates that the X-PET system is suitable for preclinical research.
机译:这项工作的目的是根据美国国家电气制造商协会(NEMA)的NU 4-2008标准,评估FLEX Triumph PET / CT扫描仪的临床前X-PET子系统的性能特征。方法:评估的性能参数包括空间分辨率,散射分数,计数损失和随机符合,灵敏度和图像质量特征。 PET检测器阵列由以48个环和180个块排列的11,520个锗酸铋晶体构成,轴向视场(FOV)为11.6 cm,内环直径为16.5 cm。使用小的(22)Na点源(直径0.25 mm)在距中心不同的径向偏移处测量了空间分辨率。使用同一光源,通过沿扫描仪的轴向FOV轴向步进来计算灵敏度。使用具有(18)F线源插入物的鼠标和大鼠大小的幻像确定散射分数和计数率性能。还执行了NEMA图像质量幻像和啮齿动物成像,以访问扫描仪的整体成像功能。结果:切向空间分辨率以半峰全宽表示,在FOV中心的2.2 mm和径向偏移2.5 cm的2.3 mm之间变化。径向空间分辨率在中心的2.0和4.4 mm(2.3 cm的径向偏移)之间变化。在FOV的中心,峰值系统绝对灵敏度为5.9%。绝对系统灵敏度为0.67 counts / s / Bq,相对总系统灵敏度为73.9%。鼠标大小的幻影的散射分数为7.9%,在0.3 MBq / mL时的峰值真实计数率为168千个计数每秒(kcps),在0.17 MBq / mL时的峰值等效噪声计数率为106 kcps。大鼠大小的体模的散射分数为21%,在0.034 MBq / mL时的真实计数率为93 kcps,在0.02 MBq / mL时的等效噪声计数率为49 kcps。图像质量体模的恢复系数范围为0.13至0.88。结论:基于NEMA NU 4-2008标准的X-PET扫描仪的性能已得到充分表征。总体性能表明,X-PET系统适用于临床前研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号