首页> 外文OA文献 >Blurred digital mammography images: an analysis of technical recall and observer detection performance
【2h】

Blurred digital mammography images: an analysis of technical recall and observer detection performance

机译:数字化乳腺摄影图像模糊:技术召回和观察者检测性能的分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective - Blurred images in full-field digital mammography are a problem in the UK Breast Screening Programme. Technical recalls may be due to blurring not being seen on lower resolution monitors used for review. This study assesses the visual detection of blurring on a 2.3-MP monitor and a 5-MP report grade monitor and proposes an observer standard for the visual detection of blurring on a 5-MP reporting grade monitor. Methods - 28 observers assessed 120 images for blurring; 20 images had no blurring present, whereas 100 images had blurring imposed through mathematical simulation at 0.2, 0.4, 0.6, 0.8 and 1.0 mm levels of motion. Technical recall rate for both monitors and angular size at each level of motion were calculated. χ2 tests were used to test whether significant differences in blurring detection existed between 2.3- and 5-MP monitors. Results - The technical recall rate for 2.3- and 5-MP monitors are 20.3% and 9.1%, respectively. The angular size for 0.2- to 1-mm motion varied from 55 to 275 arc s. The minimum amount of motion for visual detection of blurring in this study is 0.4 mm. For 0.2-mm simulated motion, there was no significant difference [χ2 (1, N = 1095) = 1.61, p = 0.20] in blurring detection between the 2.3- and 5-MP monitors. Conclusion - According to this study, monitors ≤2.3 MP are not suitable for technical review of full-field digital mammography images for the detection of blur.
机译:目的-全视野数字化乳腺摄影中的图像模糊是英国乳房筛查计划中的一个问题。技术召回可能是由于在用于检查的较低分辨率的显示器上看不到模糊。这项研究评估了在2.3-MP监视器和5-MP报告等级监视器上的视觉模糊检测,并提出了在5-MP报告级监视器上进行视觉检测的观察者标准。方法-28位观察者评估了120张图像的模糊性;通过数学模拟在运动0.2、0.4、0.6、0.8和1.0?mm时,有20张图像不存在模糊,而100张图像则具有模糊。计算了两个显示器的技术召回率和每个运动水平的角度大小。 χ2检验用于检验2.3和5-MP监视器之间在模糊检测方面是否存在显着差异。结果-2.3 MP和5 MP显示器的技术召回率分别为20.3%和9.1%。 0.2至1毫米运动的角度大小从55至275弧度变化。在本研究中,用于视觉检测模糊的最小运动量为0.4mm。对于0.2毫米的模拟运动,在2.3 MP监视器和5 MP监视器之间的模糊检测中,没有显着差异[χ2(1,N = 951095)= 1.61,p = 0.20]。结论-根据这项研究,≤2.3MP的监视器不适用于对全场乳腺X线摄影图像进行模糊检测的技术审查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号