首页> 外文OA文献 >Integration of fractal biosensor in a digital microfluidic platform
【2h】

Integration of fractal biosensor in a digital microfluidic platform

机译:分形生物传感器在数字微流控平台中的集成

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Fractal capacitive electrodes have been successfully integrated into a digital microfluidic open-platform. These electrodes perform actuation and withstand voltages up to 300V without insulation-layer breakdown. They were used to quantify the concentration levels of C-reactive protein (CRP) to determine the risk of cardiovascular disease. The capacitance increased sevenfold and stabilized in less than 5 minutes. The sensor shows a decreasing trend of capacitance readouts with the increase of concentrations. The same immunoassay was tested with untreated electrodes and showed no significant response, which suggests that immobilization was necessary. This configuration allows the electrodes to be used as biosensors.
机译:分形电容电极已成功集成到数字微流开放平台中。这些电极执行操作并承受高达300V的电压而不会发生绝缘层击穿。他们被用来量化C反应蛋白(CRP)的浓度水平,以确定心血管疾病的风险。电容增加了七倍,并在不到5分钟的时间内稳定下来。随着浓度的增加,传感器显示出电容读数的下降趋势。使用未经处理的电极测试了相同的免疫测定,未显示明显反应,这表明固定是必要的。这种配置允许电极用作生物传感器。