REGISTRASI CITRA DIGITALLY RECONTRUCTED RADIOGRAPHS TERHADAP CITRA ELECTRONIC PORTAL IMAGING DEVICES UNTUK VERIFIKASI POSISI PASIEN RADIOTERAPI

机译:在门户电子成像设备图像上对数字重建的放射线图像进行配准,以验证放射治疗患者的位置

摘要

A research on development of automatic 2D/2D image registration system for digitally reconstructed radiographs (DRR) and electronic portal imaging devices image (EPID) has been carried out to verify patient position in radiotherapy treatment. Registration is done using two methods: rigid transformation and mutual information. Implementations using rigid transformations yields X-axis average translation (2,18??0,89) mm, Y-axis (1,18??0,59) mm as well as Z-axis (2,08??0,9) mm while the average registration time which is (42.85??15.20) s. Implementations using mutual information yields X-axis average translation (2,27??0,8) mm,Y axis (1,89??0,65) mm as well as Z axis (0,94??0,59) mm while the average registration time which is (2.85??0.54) s. We can conclude that there is no significant difference in P value > 0.05 according to chi square test corresponds to verification manual, semiautomatic fusion, rigid transformations, and mutual information.
机译:为了验证患者在放射治疗中的位置,已经进行了针对数字重建射线照片(DRR)和电子门成像设备图像(EPID)的自动2D / 2D图像配准系统的开发研究。注册使用两种方法完成:刚性转换和互信息。使用刚性变换的实现产生X轴平均平移(2,18 ?? 0.89)mm,Y轴(1.18 ?? 0.59)mm和Z轴(2,08 ?? 0, 9)mm,而平均注册时间为(42.85 ?? 15.20)s。使用互信息的实现产生X轴平均平移(2,27 ?? 0,8)mm,Y轴(1,89 ?? 0,65)mm以及Z轴(0,94 ?? 0,59) mm,而平均套准时间为(2.85 ?? 0.54)s。我们可以得出结论,根据卡方检验,对应于验证手册,半自动融合,刚性变换和互信息,P值> 0.05没有显着差异。

著录项

  • 作者

    BUDIYONO ALFRED;

  • 作者单位
  • 年度 2013
  • 总页数
  • 原文格式 PDF
  • 正文语种
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号