首页> 外文OA文献 >Development of a local electromagnetic shielding for an extremity magnetic resonance imaging system
【2h】

Development of a local electromagnetic shielding for an extremity magnetic resonance imaging system

机译:用于四肢磁共振成像系统的局部电磁屏蔽的开发

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A local radio frequency (rf) shielding consisting of a Cu plate and an LC balun circuit has been developed for a compact magnetic resonance imaging (MRI) system with a 0.3 T permanent magnet. Performance of the local rf shielding was evaluated using an artificial external noise source irradiating a human subject whose hand was inserted into the rf coil of the MRI system. Power spectra of the rf signal detected through the rf coil demonstrated that the local rf shield achieved 30.1 dB external noise suppression. With the local rf shielding, a MRI of the subjectu27s hand was performed using a three-dimensional gradient-echo sequence. Anatomical structures of the subjectu27s hand were clearly visualized. It was concluded that the local rf shielding could be used for the compact MRI system instead of a rf shielded room.
机译:已经开发了由Cu板和LC不平衡变压器电路组成的局部射频(rf)屏蔽,用于带有0.3 T永磁体的紧凑型磁共振成像(MRI)系统。局部射频屏蔽的性能是通过使用人工外部噪声源辐照人类受试者的手来评估的,该人类受试者的手已插入MRI系统的射频线圈中。通过射频线圈检测到的射频信号的功率谱表明,本地射频屏蔽实现了30.1 dB的外部噪声抑制。通过局部射频屏蔽,使用三维梯度回波序列对受试者的手进行了MRI。该对象的手的解剖结构清晰可见。结论是,局部射频屏蔽可用于紧凑型MRI系统,而不是射频屏蔽室。