Interferometric phase-contrast X-ray CT imaging of VX2 rabbit cancer at 35keV X-ray energy

机译:VX2兔癌的35keV X射线能量干涉相衬X射线CT成像

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Imaging of large objects at 17.7-keV low x-ray energy causes huge x-ray exposure to the objects even using interferometric phase-contrast x-ray CT (PCCT). Thus, we tried to obtain PCCT images at high x-ray energy of 35keV and examined the image quality using a formalin-fixed VX2 rabbit cancer specimen with 15-mm in diameter. The PCCT system consisted of an asymmetrically cut silicon (220) crystal, a monolithic x-ray interferometer, a phase-shifter, an object cell and an x-ray CCD camera. The PCCT at 35 keV clearly visualized various inner structures of VX2 rabbit cancer such as necrosis, cancer, the surrounding tumor vessels, and normal liver tissue. Besides, image-contrast was not degraded significantly. These results suggest that the PCCT at 35 KeV is sufficient to clearly depict the histopathological morphology of VX2 rabbit cancer specimen.
机译:即使使用干涉相衬X射线CT(PCCT),在17.7-keV的低X射线能量下对大型物体进行成像也会导致对物体的巨大X射线曝光。因此,我们试图在35keV的高X射线能量下获得PCCT图像,并使用直径15毫米的福尔马林固定VX2兔癌标本检查了图像质量。 PCCT系统由不对称切割的硅(220)晶体,单片X射线干涉仪,移相器,物镜和X射线CCD相机组成。 35 keV的PCCT清晰可见VX2兔癌的各种内部结构,例如坏死,癌症,周围的肿瘤血管和正常肝组织。此外,图像对比度没有明显降低。这些结果表明,在35 KeV处的PCCT足以清楚地描绘VX2兔癌标本的组织病理形态。